• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AV-TEMECULA-CT RX ACCULINK CAROTID STENT SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AV-TEMECULA-CT RX ACCULINK CAROTID STENT SYSTEM Back to Search Results
Catalog Number 1011344-40
Health Effect - Clinical Codes Seizures (2063); Thrombosis (2100)
Date of Event 02/05/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).The device was not returned for evaluation.The reported patient effects of seizures and thrombosis are listed in the rx acculink instruction for use as known patient effects.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Based on the case information, a conclusive cause for the reported patient effects, and the relationship to the product, if any, cannot be determined.There is no indication of a product quality issue with respect to manufacture, design or labeling.The viatrac device is being filed under separate medwatch report.
 
Event or Problem Description
It was reported that the procedure was to treat a lesion in the internal carotid artery.On (b)(6) 2018, a 4x20x135mm viatrac plus balloon catheter was being used for pre-dilatation; however, the balloon ruptured during first inflation at 4 atmospheres.A 7-10x40mm acculink stent was implanted and the procedure was completed successfully with the patient being given plavix post-procedure.The patient had a seizure overnight and a computerized tomographic angiography confirmed thrombosis in the acculink stent.The patient is in stable condition and is going back to the cath lab on (b)(6) 2018 to have a covered stent implanted to treat the thrombosis.No additional information was provided.
 
Additional Manufacturer Narrative
(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RX ACCULINK CAROTID STENT SYSTEM
Common Device Name
CAROTID STENT SYSTEM
Manufacturer (Section D)
AV-TEMECULA-CT
abbott vascular
26531 ynez road
temecula CA 92591 4628
MDR Report Key7265234
Report Number2024168-2018-01026
Device Sequence Number1475135
Product Code NIM
UDI-Device Identifier08717648076398
UDI-Public08717648076398
Combination Product (Y/N)N
Initial Reporter StateKY
Initial Reporter CountryUS
PMA/510(K) Number
P040012
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Health Professional
Remedial Action Other
Type of Report Initial,Followup
Report Date (Section B) 04/02/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date11/30/2018
Device Catalogue Number1011344-40
Device Lot Number6120861
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 02/05/2018
Supplement Date Received by Manufacturer03/28/2018
Initial Report FDA Received Date02/12/2018
Supplement Report FDA Received Date04/02/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-