| Model Number |
8253001RF |
| Medical Device Problem Code |
Device Stops Intermittently (1599)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
01/18/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The product analysis indicates the mainframe came in with a white line across the left side of the touchscreen display upon boot-up.The reported issue could not be verified.The lcd display was replaced.The mainframe was placed into burn-in for four hours.The device was tested and passed to specifications.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
The customer reported that the nim system was intermittently not working.The case was completed with a backup device.There was no patient impact.This is regulatory report 1 of 2 (nim mainframe).
|
| |
|