| Model Number |
8253200 |
| Medical Device Problem Code |
Device Stops Intermittently (1599)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
01/18/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The product analysis indicates the patient interface came in with missing spare fuses.The reported issue could not be duplicated.As a precaution, the p/i board was replaced.The cable failed triboelectric immunity test and was replaced.The spare fuses were also replaced.The patient interface was successfully tested to manufacturing specifications.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
The customer reported that the nim system was intermittently not working.The case was completed with a backup device.There was no patient impact.This is regulatory report 2 of 2 (patient interface).
|
| |
|