Model Number 8253200
Medical Device Problem Code Device Stops Intermittently (1599)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 01/18/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The product analysis indicates the patient interface came in with missing spare fuses.The reported issue could not be duplicated.As a precaution, the p/i board was replaced.The cable failed triboelectric immunity test and was replaced.The spare fuses were also replaced.The patient interface was successfully tested to manufacturing specifications.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
The customer reported that the nim system was intermittently not working.The case was completed with a backup device.There was no patient impact.This is regulatory report 2 of 2 (patient interface).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NIM® 3.0 INTERFACE
Common Device Name
STIMULATOR, NERVE
Manufacturer (Section D)
MEDTRONIC XOMED INC.
6743 southpoint dr n
jacksonville FL 32216
MDR Report Key7266040
Report Number1045254-2018-00052
Device Sequence Number19288432
Product Code ETN
UDI-Device Identifier00643169782501
UDI-Public00643169782501
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K083124
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional,user faci
Initial Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date (Section B) 02/13/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/20/2013
Device Model Number8253200
Device Catalogue Number8253200
Device Lot Number206555444
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/24/2018
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/18/2018
Initial Report FDA Received Date02/13/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-