ANIMAS CORPORATION ANIMAS VIBE; INSULIN INFUSION PUMP Back to Search Results
Medical Device Problem Codes Premature Discharge of Battery (1057); Structural Problem (2506)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The pump has not been returned to animas for evaluation.If the device is returned, an evaluation shall be completed and a supplemental report will be filed.No conclusions can be made at this time.
 
Event or Problem Description
On (b)(6) 2018, the reporter contacted animas, alleging that the pump had shorter than expected battery life.This complaint is being reported because the issue may result in a long-term cessation of insulin delivery if the user is unable to resolve the alarm.There was no serious injury associated with the complaint.
 
Additional Manufacturer Narrative
Device evaluation: the device has been returned and evaluated by product analysis on 15-jun-2018 with the following findings: the black box showed partially discharged batteries were installed leading to replace battery alarms and low battery warnings.Unrelated to the original complaint the battery compartment was cracked above and below the bumper pad.The battery compartment threads were stripped.The battery cap was not returned for inspection.No moisture/corrosion was observed in the battery compartment.A test battery cap was able to attach to the pump and power it on.Current draws measured within specifications.The pump cover was; no evidence of moisture or loose components was found inside the pump.A "battery life" issue was not duplicated during testing.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ANIMAS VIBE
Common Device Name
INSULIN INFUSION PUMP
Manufacturer (Section D)
ANIMAS CORPORATION
200 lawrence dr
west chester PA 19380 3428
MDR Report Key7266073
Report Number2531779-2018-03306
Device Sequence Number1655859
Product Code OYC
UDI-Device Identifier10840406100372
UDI-Public10840406100372
Combination Product (Y/N)Y
Initial Reporter StateNJ
Initial Reporter CountryUS
PMA/510(K) Number
P130007
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer
Initial Reporter Occupation Other Caregivers
Type of Report Initial,Followup
Report Date (Section B) 01/29/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Other Device ID Number1-MCBC-3762
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/06/2018
Was the Report Sent to FDA? Yes
Device Age15 MO
Initial Date Received by Manufacturer 01/29/2018
Supplement Date Received by Manufacturer01/29/2018
Initial Report FDA Received Date02/13/2018
Supplement Report FDA Received Date06/22/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age16 YR
Patient Weight136
-
-