MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS INC 1818910 PFC* MOD PLUS TORQUE DRIVER; KNEE INSTRUMENT : SCREWDRIVERS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS INC 1818910 PFC* MOD PLUS TORQUE DRIVER; KNEE INSTRUMENT : SCREWDRIVERS Back to Search Results
Catalog Number 860284
Medical Device Problem Codes Break (1069); Appropriate Device Problem Term/Code Not Available (3191)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Product complaint # (b)(4).
 
Event or Problem Description
860284.Lot unknown -pfc mod plus torque driver didn¿t appear to click at maximum torque causing the wobble bits to fracture.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Additional Manufacturer Narrative
Product complaint #: (b)(4).Investigation summary: examination and functional testing of the returned device was unable to confirm the reported event.Depuy considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Common Device Name
KNEE INSTRUMENT : SCREWDRIVERS
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC 1818910
700 orthopaedic drive
warsaw IN 46582 0988
MDR Report Key7266088
Report Number1818910-2018-53304
Device Sequence Number2117182
Product Code HWR
UDI-Device Identifier10603295215868
UDI-Public10603295215868
Combination Product (Y/N)N
Initial Reporter CountryUK
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Health Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 01/29/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number860284
Device Lot NumberSW0614
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/22/2018
Initial Date Received by Manufacturer 01/29/2018
Supplement Date Received by Manufacturer03/16/2018
09/19/2018
Initial Report FDA Received Date02/13/2018
Supplement Report FDA Received Date04/02/2018
10/19/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-