ALLERGAN (COSTA RICA) STYLE 133PLUS MV TISSUE EXPANDER WITH SUTURE TAB; EXPANDER, SKIN, INFLATABLE Back to Search Results
Catalog Number 133P-MV-15-T
Medical Device Problem Code Fluid/Blood Leak (1250)
Health Effect - Clinical Code Failure of Implant (1924)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
A review of the device history record has been completed.No deviations or non-conformances noted.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Device labeling: the following is a list of potential adverse events that may occur with breast implant surgery.The risks include: implant deflation/leakage, additional surgery, capsular contracture, infection, toxic shock syndrome, necrosis, hematoma, seroma, extrusion, breast pain, changes in nipple sensation, changes in breast sensation, dissatisfaction with cosmetic results (wrinkling, folding, displacement, asymmetry, palpability, visibility, ptosis, sloshing), calcific deposits, irritation/inflammation, delayed wound healing, hypertrophic scarring, breast tissue atrophy/chest wall deformity, difficulty/inability in breast feeding, and inability to adequately visualize breast lesions with mammography.In addition to these potential adverse events, there have been concerns with certain systemic diseases.Deflation - patients should be advised that the tissue expanders may deflate, and require replacement surgery.Deflation occurs when saline leaks through a damaged injection site or a damaged tissue expander envelope.
 
Event or Problem Description
Health professional reported a left side deflation of a tissue expander.Device has been explanted and replaced with permanent implant.
 
Additional Manufacturer Narrative
Device evaluation: visual analysis of the returned device identified: opening, fluid, and yellow particles.Leak test was performed and found opening on posterior side.A microscopic analysis was performed which identified a striated edge opening on posterior side consistent with use of a surgical tool, and sharp edge opening assessed as unidentified (tear) opening.A dimensional measurement in the shell was performed which identified the thickness within specification.Based on the device analysis the final assessment is: a striated edge opening on posterior side due to surgical damage.A sharp opening on posterior due to an unidentified (tear) opening.
 
Event or Problem Description
Health professional reported a left side deflation of a tissue expander.Device has been explanted and replaced with permanent implant.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STYLE 133PLUS MV TISSUE EXPANDER WITH SUTURE TAB
Common Device Name
EXPANDER, SKIN, INFLATABLE
Manufacturer (Section D)
ALLERGAN (COSTA RICA)
900 parkway global park
zona franca
CS 
MDR Report Key7266303
Report Number9617229-2018-00793
Device Sequence Number1455294
Product Code LCJ
UDI-Device Identifier10888628010093
UDI-Public10888628010093
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K143354
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Device Explanted Year2017
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 03/30/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/25/2021
Device Catalogue Number133P-MV-15-T
Device Lot Number2917473
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/28/2018
Initial Date Received by Manufacturer 01/24/2018
Supplement Date Received by Manufacturer03/08/2018
Initial Report FDA Received Date02/13/2018
Supplement Report FDA Received Date03/30/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
-
-