BOSTON SCIENTIFIC - SPENCER ADVANIX¿ BILIARY; CATHETER, BILIARY, DIAGNOSTIC Back to Search Results
Model Number M00532960
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 12/21/2017
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).A visual evaluation of the returned device found that the stent is severely damaged, it is kinked in several locations and the proximal section is stretched and broken.Based on the product analysis, most likely some anatomical/procedural factors were encountered during the procedure which may have limited the overall performance of the device.A severe stricture due to patient anatomy or customer use/manipulation/maneuvering can lead to difficult to advance the device through the patient anatomy, continuous attempts to advance the device can cause damages to the stent.Likely an additional device was used for retrieve the stent, this additional device could have contributed with the damages found on the device including the proximal section stretched and broken.Therefore, ¿operational context¿ is selected as the most probable root cause for the complaint.A dhr (device history record) review was performed and no deviation was found.A search of the complaint database revealed that no similar complaints exist for the specified lot.
 
Event or Problem Description
It was reported to boston scientific corporation that an advanix¿ biliary stent was used during an endoscopic retrograde cholangiopancreatography (ercp) procedure performed in the common bile duct on (b)(6) 2017.According to the complainant, during the procedure, the stent was difficult to advance into the common bile duct.The procedure was completed with another advanix¿ biliary stent.There were no patient complications reported as a result of this event.The patient's condition at the conclusion of the procedure was reported to be "fine".Investigation results revealed the stent was broken, therefore this event has been deemed an mdr reportable event.Please see block h10 for full investigation details.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ADVANIX¿ BILIARY
Common Device Name
CATHETER, BILIARY, DIAGNOSTIC
Manufacturer (Section D)
BOSTON SCIENTIFIC - SPENCER
780 brookside drive
spencer IN 47460
MDR Report Key7266888
Report Number3005099803-2018-00402
Device Sequence Number1316724
Product Code FGE
UDI-Device Identifier08714729786870
UDI-Public08714729786870
Combination Product (Y/N)N
Initial Reporter CountryKR
PMA/510(K) Number
K101314
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 12/21/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/10/2019
Device Model NumberM00532960
Device Catalogue Number3296
Device Lot Number0020991923
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/23/2018
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/23/2018
Initial Report FDA Received Date02/13/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-