• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER ORTHOPAEDICS-MAHWAH X3 TRIATHLON CS INSERT #5 9MM; PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, CO

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ORTHOPAEDICS-MAHWAH X3 TRIATHLON CS INSERT #5 9MM; PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, CO Back to Search Results
Catalog Number 5531G509
Device Problems Break (1069); Fracture (1260); Scratched Material (3020)
Patient Problem Injury (2348)
Event Date 01/15/2018
Event Type  Injury  
Manufacturer Narrative
Review of the device history records indicate devices were manufactured and accepted into final stock with no reported discrepancies.There have been no other events for the lot referenced.A supplemental report will be submitted upon completion of the investigation.
 
Event Description
Inlay was dorsal broken.
 
Event Description
Inlay was dorsal broken.
 
Manufacturer Narrative
An event regarding crack/fracture involving a triathlon insert was reported.The event was confirmed based on the returned insert.Conclusion: visual inspection confirmed that the explanted insert component was returned damaged, with material loss in one of the posterior compartments on the articulating surface.Material analysis concluded: [.] a localized area of material loss was observed along the articulating surface; posterior edge, likely resulting from the cyclic edge loading of the femoral component.There was no evidence of manufacturing or material defects on the returned ultra-high molecular weight polyethylene (uhmwpe) tibial insert.The cause of the event cannot be determined as insufficient information was provided.Further information such as x-rays and medical records are required to complete the investigation for determining a root cause.No further investigation is required at this time.If additional information becomes available to indicate further evaluation is warranted, this record will be reopened.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
X3 TRIATHLON CS INSERT #5 9MM
Type of Device
PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, CO
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
MDR Report Key7266956
MDR Text Key99924895
Report Number0002249697-2018-00399
Device Sequence Number1
Product Code MBH
Combination Product (y/n)N
PMA/PMN Number
K141056
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,other
Type of Report Initial,Followup
Report Date 03/26/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date06/30/2020
Device Catalogue Number5531G509
Device Lot NumberLEK377
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/23/2018
Initial Date Manufacturer Received 01/16/2018
Initial Date FDA Received02/13/2018
Supplement Dates Manufacturer Received02/26/2018
Supplement Dates FDA Received03/26/2018
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Other; Required Intervention;
-
-