STRYKER ORTHOPAEDICS-MAHWAH X3 TRIATHLON CS INSERT #5 9MM; PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, CO
|
Back to Search Results |
|
Catalog Number 5531G509 |
Device Problems
Break (1069); Fracture (1260); Scratched Material (3020)
|
Patient Problem
Injury (2348)
|
Event Date 01/15/2018 |
Event Type
Injury
|
Manufacturer Narrative
|
Review of the device history records indicate devices were manufactured and accepted into final stock with no reported discrepancies.There have been no other events for the lot referenced.A supplemental report will be submitted upon completion of the investigation.
|
|
Event Description
|
Inlay was dorsal broken.
|
|
Event Description
|
Inlay was dorsal broken.
|
|
Manufacturer Narrative
|
An event regarding crack/fracture involving a triathlon insert was reported.The event was confirmed based on the returned insert.Conclusion: visual inspection confirmed that the explanted insert component was returned damaged, with material loss in one of the posterior compartments on the articulating surface.Material analysis concluded: [.] a localized area of material loss was observed along the articulating surface; posterior edge, likely resulting from the cyclic edge loading of the femoral component.There was no evidence of manufacturing or material defects on the returned ultra-high molecular weight polyethylene (uhmwpe) tibial insert.The cause of the event cannot be determined as insufficient information was provided.Further information such as x-rays and medical records are required to complete the investigation for determining a root cause.No further investigation is required at this time.If additional information becomes available to indicate further evaluation is warranted, this record will be reopened.
|
|
Search Alerts/Recalls
|
|
|