Catalog Number 1125250-33
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 01/29/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The patient and device codes were coded by the manufacturer.(b)(4).During processing of this complaint, attempts were made to obtain complete event, patient and device information.The device was returned for analysis.The reported stent dislodgement was confirmed.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no similar incidents from this lot.The investigation determined the reported difficulties appear to be related to circumstances of the procedure.There is no indication of a product quality issue with respect to manufacture, design or labeling.
 
Event or Problem Description
It was reported that the procedure was to treat a heavily calcified de novo distal right coronary artery that did not have any tortuosity.A 2.5 x 33 mm xience alpine stent delivery system (sds) was used; however, it was noted under fluoroscopy that the stent was missing from the balloon.The stent was dislodged inside an unspecified guiding catheter.The sds was then pulled back and the sds balloon was inflated.The sds and the guiding catheter were removed with the stent inside the guiding catheter.A new unspecified guiding catheter was then advanced and a 2.5 x 23 mm xience alpine sds was used; however, the sds failed to cross the lesion due to the anatomy.Therefore, the device was replaced with a 2.5 x 12 mm and a 2.5 x 18 mm xience alpine sds, and post-dilatation was done with an unspecified balloon catheter to successfully complete the procedure.There was no clinically significant delay in the procedure and no adverse patient effects.No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
XIENCE ALPINE EVEROLIMUS ELUTING CORONARY STENT SYSTEM
Common Device Name
DRUG ELUTING CORONARY STENT SYSTEM
Manufacturer (Section D)
AV-TEMECULA-CT
abbott vascular
26531 ynez road
temecula CA 92591 4628
MDR Report Key7267549
Report Number2024168-2018-01050
Device Sequence Number13816996
Product Code NIQ
UDI-Device Identifier08717648199202
UDI-Public08717648199202
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
PMA/510(K) Number
P110019
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Health Professional
Remedial Action Other
Type of Report Initial
Report Date (Section B) 02/13/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/23/2020
Device Catalogue Number1125250-33
Device Lot Number7111741
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/01/2018
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/29/2018
Initial Report FDA Received Date02/13/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age43 YR
Patient Weight87
-
-