| Model Number |
TRUETRACK |
| Medical Device Problem Codes |
High Test Results (2457); Device Displays Incorrect Message (2591)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
01/23/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Returned meter evaluated with defect found.Returned test strips evaluated with no defect found.Qc investigation on 2/2/2018 resulted in defect found; returned meter code key damage/dirty.Reported defect not reproduced with returned meter and test strip.Final root cause investigation has been completed: pcb contamination on the code key connector j2 caused by user.Test strip udi# (b)(4).Manufacturer contacted customer on (b)(6) 2018 in a follow-up call in order to ensure that the replacement products resolved the initial concern; customer stated that the replacement product is working to their satisfaction and they have not had any medical intervention since the last call.
|
| |
|
Event or Problem Description
|
|
Consumer reported complaint for high blood glucose results.The customer is concerned with results obtained of 423 and 559 mg/dl.The expected fasting blood glucose test result range is 75 to 110 mg/dl.The customer feels well and did not report symptoms.Medical attention is not reported as a result of the actual blood glucose results.The product is stored by customer according to specification in the living room.During the call on (b)(6) 2018, a back to back blood test was not performed by the customer.The test strip lot manufacturer's expiration date is 07/13/2019 and open vial date is (b)(6) 2018.The meter memory was reviewed for previous test result history: (date/time not stored correctly): (b)(6).When attempting to do the back to back testing the meter read code (---) and customer could not test.
|
| |
|
Search Alerts/Recalls
|