The reported unused devices were returned for evaluation in original unopened packaging.Label verification and visual inspection were performed.A hole was found in the poly material, adjacent to the device's proximal electrode, of both packages.These defects resulted in a sterile breach and were likely caused by the device after the sealing process.The manufacturing documents from the device history record have been reviewed with special attention to the manufacturing and inspection of this product.The products released for distribution were found to have met all specifications prior to shipment.Of the lot containing 2400 units, no prior complaints have been received.A two-year review of complaint history revealed seven similar prior complaints.During this same time frame, 251,600 devices have been manufactured and shipped worldwide, making the rate of occurrence of this failure 0.003 percent.This reported packaging issue was obvious to the distributor, prompting the return of the device for evaluation.There was no patient involvement.As with all medical devices, examination of the product occurs multiple times prior to use.Good clinical practice would include examination and verification of the original packaging and its labeling to ensure both are intact.The instructions for use (ifu) provides the following warning.-if packaging has been opened/damaged or altered, do not use the product and contact the manufacturer immediately.This issue will continue to be monitored through the complaint system to assure patient safety.
|