• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: NEODENT - JJGC S.A. CM DRIVE ACQUA IMPLANT, TITANIUM, 5.0X18; ENDOSSEOUS DENTAL IMPLANT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

NEODENT - JJGC S.A. CM DRIVE ACQUA IMPLANT, TITANIUM, 5.0X18; ENDOSSEOUS DENTAL IMPLANT Back to Search Results
Catalog Number 140.898
Medical Device Problem Code Failure to Osseointegrate (1863)
Health Effect - Clinical Codes Pain (1994); Tissue Damage (2104)
Date of Event 01/04/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Exemption number: e2015015.Instradent usa, inc is submitting the report on behalf of neodent - jjgc.
 
Event or Problem Description
The clinician reports that 5 months after the dental implant and abutment were placed in ada site 15 in type i bone, the implant had not yet achieved osseointegration.A bone graft was placed.Clinician reports discomfort which went away after the implant's removal.The product will be forwarded to the manufacturer for investigation.There were no reported post-operative complications.
 
Additional Manufacturer Narrative
Exemption number: e2015015.Instradent usa, inc is submitting the report on behalf of neodent - jjgc.The dentist also reported that immediate load and immediate implant were performed.
 
Event or Problem Description
The clinician reports that 5 months after the dental implant and abutment were placed in ada site 15 in type i bone, the implant had not yet achieved osseointegration.A bone graft was placed.Clinician reports discomfort which went away after the implant's removal.The product will be forwarded to the manufacturer for investigation.There were no reported post-operative complications.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CM DRIVE ACQUA IMPLANT, TITANIUM, 5.0X18
Common Device Name
ENDOSSEOUS DENTAL IMPLANT
Manufacturer (Section D)
NEODENT - JJGC S.A.
av. juscelino kubitschek de
oliveira, 3291 cic
curitiba PR 81270 -200
BR  81270-200
MDR Report Key7268215
Report Number3008261720-2018-00689
Device Sequence Number1455216
Product Code DZE
UDI-Device Identifier07899878018087
UDI-Public07899878018087
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K150182
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Device Explanted Year2018
Reporter Type Manufacturer
Report Source distributor,foreign
Initial Reporter Occupation Dentist
Type of Report Initial,Followup
Report Date (Section B) 02/13/2018
Report Date (Section F) 01/18/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Dentist
Device Expiration Date09/27/2018
Device Catalogue Number140.898
Device Lot Number69856I
Other Device ID Number07899878018087
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/10/2018
Was the Report Sent to FDA? No
Distributor Facility Aware Date05/02/2018
Device Age1 MO
Event Location Other
Date Report to Manufacturer01/18/2018
Initial Date Received by Manufacturer 01/18/2018
Supplement Date Received by Manufacturer04/10/2018
Initial Report FDA Received Date02/13/2018
Supplement Report FDA Received Date05/02/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
ABUTMENT; ABUTMENT; ABUTMENT
Outcome Attributed to Adverse Event Required Intervention;
-
-