| Model Number |
XX-5202J |
| Medical Device Problem Codes |
Detachment Of Device Component (1104); Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
01/31/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
According to the reporter, during the laparoscopic hysterectomy procedure, the device needle detached easily and fell into the cavity.The needle was soon retrieved.They then used another device to resolve the issue.There was no patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
Evaluation summary post market vigilance (pmv) led an evaluation of one device.The visual inspection of the samples had acceptable results.A review of the device history record indicates this product was released meeting all quality release specifications at the time of manufacture.However, an assembly error was identified during product analysis.The root cause of the reported condition was an improperly made code along with an improper execution of in-process controls for material assignment.Should new information become available, the file will be re-opened and the investigation summary amended as appropriate.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
According to the reporter, during the laparoscopic hysterectomy procedure, the device needle detached easily and fell into the cavity.The needle was soon retrieved.An x-ray was not performed for the express purpose of locating the component.They then used another device to resolve the issue.There was no patient injury.
|
| |
|