• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: LIVANOVA CANADA CORP. MITROFLOW AORTIC PERICARDIAL HEART VALVE WITH PRT; TISSUE HEART VALVES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

LIVANOVA CANADA CORP. MITROFLOW AORTIC PERICARDIAL HEART VALVE WITH PRT; TISSUE HEART VALVES Back to Search Results
Model Number DLA25
Medical Device Problem Code Calcified (1077)
Health Effect - Clinical Codes Corneal Pannus (1447); Aortic Valve Stenosis (1717); Thrombus (2101)
Date of Event 01/10/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Further information is being requested by the manufacturer.Visual examination was performed on the returned device on (b)(6) 2018.The returned valve was received with a piece detached and traces of pannus on the sewing cuff, in particular on the outflow side.Calcifications on the leaflet were visible.Further evaluation is being performed by the manufacturer and will be reported in the follow up report.
 
Event or Problem Description
Manufacturer was notified that patient who had mitroflow dla valve implanted on (b)(6) 2016 underwent aortic valve replacement on (b)(6) 2018 and the device was replaced with mechanical valve.
 
Event or Problem Description
Manufacturer was notified that patient who had mitroflow dla 25 mm valve implanted on (b)(6) 2016 underwent aortic valve replacement on (b)(6) 2018, due to stenosis.The valve leaflets were reportedly thickened and rigid.The device was replaced with a st.Jude regent 23 mm mechanical valve, and the patient outcome was good.
 
Additional Manufacturer Narrative
Radiographic review of the returned device confirmed the presence of intrinsic calcification in all of the leaflets.Hystopathologic analysis confirmed the presence of pannus over the free edges of both leaflets surrounding one of the valve posts, which contributed to the observed stenosis of the valve.All leaflets were stiffened due to intrinsic and vegetating calcifications.Small thrombus depositions were visible on the inflow surfaces.Gram staining showed some focal gram positive bacteria.The manufacturing and material records for the mitroflow bioprosthetic pericardial heart valve, model # dla25, s/n # (b)(4), as they pertain to the reported event, were pulled and reviewed by quality engineering at livanova (b)(4) corp.The results confirmed that this valve satisfied all material, visual, and performance standards required for a dla25 mitroflow aortic pericardial heart valve at the time of manufacture and release.As reported among the contraindications in the mitroflow dla instructions for use, patients "who are 55 years of age or less may have an increased tendency toward calcification of valvular bioprostheses." as such, the patient's age ((b)(6)) was a risk factor for early structural valve deterioration, and a contraindication for use of the mitroflow dla, and thus may have contributed to the reported failure of this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MITROFLOW AORTIC PERICARDIAL HEART VALVE WITH PRT
Common Device Name
TISSUE HEART VALVES
Manufacturer (Section D)
LIVANOVA CANADA CORP.
5005 north fraser way
burnaby, bc V5J5M 1
CA  V5J5M1
MDR Report Key7268657
Report Number3004478276-2018-00127
Device Sequence Number11126635
Product Code LWR
UDI-Device Identifier00896208000184
UDI-Public(01)00896208000184(240)DLA25(17)201113
Combination Product (Y/N)N
PMA/510(K) Number
P060038
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2016
Device Explanted Year2018
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Health Professional
Type of Report Initial,Followup
Report Date (Section B) 03/09/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Physician
Device Expiration Date11/13/2020
Device Model NumberDLA25
Device Catalogue NumberDLA25
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/06/2018
Initial Date Received by Manufacturer 01/14/2018
Supplement Date Received by Manufacturer02/11/2018
Initial Report FDA Received Date02/13/2018
Supplement Report FDA Received Date03/09/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age45 YR
Patient Weight100
-
-