| Catalog Number |
PXB35-07-17-080 |
| Medical Device Problem Code |
Fracture (1260)
|
| Health Effect - Clinical Codes |
Pain (1994); Patient Problem/Medical Problem (2688)
|
| Date of Event |
01/15/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
The physician treated a 15mm moderately calcified lesion in the proximal celiac artery with a visi-pro balloon- expandable stent.The artery had a diameter of 7mm and had little tortouosity.The visipro was prepped as per the ifu with no issues identified.The procedure used a 7 fr sheath and a non-medtronic wire.014" x 300 cm.No resistance was encountered advancing the device and no excessive force was used.Two weeks after the procedure, patient complained of pain and a follow cannulation of the artery was performed.During this follow-up visit, it was noticed that the implanted stent had fractured.The component embolized in the vessel.The broken part of the device was explanted using a snare.The physician opted for no further treatment.The patient complained of abdominal pain and was referred to a gi physician.No further complications reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Evaluation summary: a small fractured segment of the visi-pro balloon-expandable biliary stent crimped onto a guidewire was received for evaluation.No cine images were received.Approximately 8mm of the stent was returned crimped onto a guidewire.Sanguine residue and biological debris was noted in the stent strut cells.The returned stent structure exhibited stent cell elongation and lateral compression.If information is provided in the future, a supplemental report will be issued.
|
| |
|