• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDELA LLC BREASTPUMP PNSA STARTER; PUMP, BREAST, POWERED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDELA LLC BREASTPUMP PNSA STARTER; PUMP, BREAST, POWERED Back to Search Results
Model Number 57081
Health Effect - Clinical Code Unspecified Infection (1930)
Date of Event 02/07/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The customer was sent a replacement pump and return of her original pump was requested for testing/evaluation.Attempts to get additional information from the customer are ongoing.Based on the results of ca11-001, it cannot be definitively concluded that the pump caused or contributed to the customer¿s mastitis.The estimated incidence of mastitis in lactating women, whether using a breast pump or not, according to published clinical literature can be as high as 33%.In fact, clinical guidelines suggest the use of a breast pump to facilitate withdrawal of breast milk during bouts of mastitis.The complaint rate of mastitis across all reported failures, across all medela breast pumps, is 0.008% for the period of january 2013 to august 2017."mastitis is usually a benign, self-limiting infection with few consequences for the suckling infant.The risk of mastitis is higher among women who have breastfed previously, especially those with a history or mastitis." riordan and wambach, 4th ed.P.294: breastfeeding and human lactation.Mastitis requires prompt medical attention for the mother for pain relief and prescription antibiotics to avoid progression to overwhelming sepsis.
 
Event or Problem Description
On (b)(6) 2018, the customer alleged to medela llc that her pump in style breast pump had low suction and she had mastitis, for which she was prescribed an antibiotic.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BREASTPUMP PNSA STARTER
Common Device Name
PUMP, BREAST, POWERED
Manufacturer (Section D)
MEDELA LLC
1101 corporate dr
mchenry IL 60050
MDR Report Key7269472
Report Number1419937-2018-00042
Device Sequence Number14021895
Product Code HGX
UDI-Device Identifier00020451570815
UDI-Public0100020451570815
Combination Product (Y/N)N
Initial Reporter StateOK
PMA/510(K) Number
K031614
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 02/13/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Model Number57081
Device Catalogue Number57081
Was Device Available for Evaluation? No
Distributor Facility Aware Date02/07/2018
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/07/2018
Initial Report FDA Received Date02/14/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
-
-