Clinical evaluation performed by medical affairs on 02 february 2018: hip revision surgery occurred 4 years after primary implantation.Radiographic images provided show the presence of radiolucent lines in gruen zones 1 and 2.Aseptic loosening is a possible, literature described adverse event after primary tha and causes are often unknown.In this case, the reason of failure cannot be determined.Clinical evaluation performed on 13 february 2018 by r&d product manager: the stem was analyzed.Some signs were evident on the neck, due to the revision surgery to remove the stem.The ha coating was totally absorbed.There are no signs that indicate a failure of the device and from the received part it is not possible to determine the root cause of the event.Batch review performed on 13 february 2018, lot 120920: (b)(4) items manufactured and released on 30 may 2012.Expiration date: 2017-04-30 no anomalies found related to the problem.To date, (b)(4) items of the same lot have been already sold without any similar reported event.
|