• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: RESPIRONICS, INC TRILOGY 200; VENTILATOR, CONTINUOUS, FACILITY USE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

RESPIRONICS, INC TRILOGY 200; VENTILATOR, CONTINUOUS, FACILITY USE Back to Search Results
Model Number 1040005
Medical Device Problem Code Failure To Run On AC/DC (1001)
Health Effect - Clinical Code No Patient Involvement (2645)
Date of Event 01/29/2018
Type of Reportable Event Malfunction
Event or Problem Description
A ventilator was returned to the manufacturer for routine preventive maintenance.There was no allegation of device malfunction.The device was not in patient use.During the evaluation of the device at the manufacturer's service center, the device failed to power on.The device had evidence of physical damage.The device's ac inlet connector was replaced to address the issue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRILOGY 200
Common Device Name
VENTILATOR, CONTINUOUS, FACILITY USE
Manufacturer (Section D)
RESPIRONICS, INC
1001 murry ridge lane
murrysville PA 15668
MDR Report Key7269649
Report Number2518422-2018-00366
Device Sequence Number1054704
Product Code CBK
UDI-Device Identifier00606959026223
UDI-Public00606959026223
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K093416
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Service and Testing Personnel
Type of Report Initial
Report Date (Section B) 01/29/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number1040005
Device Catalogue Number1040005
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/16/2017
Was the Report Sent to FDA? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 01/29/2018
Initial Report FDA Received Date02/14/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-