• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ETHICON ENDO-SURGERY, LLC. ENDOPATH XCEL BLADELESS TROCAR; LAPAROSCOPE, GENERAL AND PLASTIC SURGERY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ETHICON ENDO-SURGERY, LLC. ENDOPATH XCEL BLADELESS TROCAR; LAPAROSCOPE, GENERAL AND PLASTIC SURGERY Back to Search Results
Catalog Number B5ST
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 01/22/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Batch # unk.Attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.The device history records were reviewed and the manufacturing criteria were met prior to the release of this lot.
 
Event or Problem Description
It was reported that during a laparoscopic myomectomy, the tip of the outer sleeve was bend after a myoma borer was inserted.Then, the tip of the outer sleeve was broken off when the device was removed from the patient.The device was used on the uterus.No pieces were left inside the patient.Another device was used to complete the case.There were no adverse consequences to the patient.
 
Additional Manufacturer Narrative
(b)(4).Batch # p9430u, p93y7x.Device analysis: the analysis results found that the b5st instrument was received with the sleeve melted.In addition, the tyvek was returned along with the instrument.This type of damage is consistent with the one produced by an energized device.In order to prevent this kind of damage please avoid the contact with laser, electrosurgical or ultrasonic devices.The batch history record was reviewed and no defects, ncr¿s or protocols related to the complaint, were found during the manufacturing process.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDOPATH XCEL BLADELESS TROCAR
Common Device Name
LAPAROSCOPE, GENERAL AND PLASTIC SURGERY
Manufacturer (Section D)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
MDR Report Key7270437
Report Number3005075853-2018-07992
Device Sequence Number142872
Product Code GCJ
UDI-Device Identifier20705036001239
UDI-Public20705036001239
Combination Product (Y/N)N
PMA/510(K) Number
K032676
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 01/22/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/31/2022
Device Catalogue NumberB5ST
Device Lot NumberP4T71U
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/14/2018
Initial Date Received by Manufacturer 01/22/2018
Supplement Date Received by Manufacturer02/15/2018
Initial Report FDA Received Date02/14/2018
Supplement Report FDA Received Date03/02/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-