| Catalog Number |
03.632.400 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
01/22/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Device was used for treatment, not diagnosis.Patient id: (b)(6).Device is an instrument and is not implanted/explanted.Date returned to manufacturer.The device was received, the investigation could not be completed, and no conclusion could be drawn, as product is entering the complaint system.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
It was reported the shaft of a straight tip t25 driver standard broke while trying to remove a set screw during surgery.All pieces were recovered and there was no patient consequence.All available details have been provided regarding this event.Concomitant device report: [unknown set screw] (part #: unknown, lot #: unknown, quantity: 1).This complaint is for one (1) device.This report is 1 of 1 for (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Explant date is not applicable since this is an instrument.A device history record review was performed for the subject device: part no.: 03.632.400, lot no.: 8616612, manufacturing location: (b)(4), release to warehouse date: 06.Dec.2013.No ncrs were generated during production.Review of the device history record(s) showed that there were no issues during the manufacture of the product that would contribute to this complaint condition.A product development investigation was performed for the subject device: it was reported the shaft of a straight tip t25 driver standard broke while trying to remove a set screw during surgery.All pieces were recovered and there was no patient consequence.All available details have been provided regarding this event.The straight tip t25 driver (03.632.400) is one of four t25 driver shafts intended to be utilized in final locking tap tightening for the matrix system (03.632.002, 03.632.072, 03.632.400 and 03.632.401).Proper technique includes the use of a 10nm torque limiting ratchet handle (03.620.061) and a countertorque (03.632.049).The following caution is noted: ¿never use fixed or ratcheting t-handle screwdriver for this technique.If the torque limiting attachment is not used when using any of the stardrive shaft options breakage of the driver may occur and could potentially harm the patient.¿ (technique addition - final tightening) the returned 4+ year old reusable driver (manufactured in dec 2013) was examined and the complaint condition was able to be confirmed as the driver tip was found to be broken.The edge break is oblique but jagged.The material surface at the fracture site appears homogeneous when viewed under 5x magnification.No distal sheared off fragment(s) were returned.Whether this complaint can be replicated at customer quality (cq) via functional test is not applicable for this complaint condition as the returned device is already broken.No definitive root cause was able to be determined; the failure mode is consistent with incorrect technique/application of excessive force.A device history review, including material and hardness reviews, was performed for the returned instrument¿s lot number and no ncrs, mrrs or complaint-related issues were identified with the lot number which may have contributed to the complaint condition.Relevant drawings for the returned device were reviewed (both current revision and from the time of manufacture): top-level (03_632_400) and t25 tip profile (es0675).The design, materials and finishing processes were found to be appropriate for the intended use of this device.No dimensional analysis was possible due to post-manufacturing damage (oblique jagged break).Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
Concomitant medical products: 1 qty, p/n 04.632.000, l/n unknown.
|
| |
|