WRIGHTS LANE SYNTHES USA PRODUCTS LLC TI MATRIX TOP LOADING POLYAXIAL HEAD; ORTHOSIS, SPINAL PEDICLE FIXATION, FOR DDD
|
Back to Search Results |
|
| Model Number |
04.632.001 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Patient¿s height reported as (b)(6) cm.Date of postoperative matrix top-loading polyaxial head popped off is unknown.Additional device product codes: mnh, mni, kwq, kwp.Implanted on an unknown date approximately one year ago.Complainant part is not expected to be returned for manufacturer review/investigation.Without a lot number the device history records review could not be completed.The investigation could not be completed; no conclusion could be drawn, as no product was received.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
It was reported that a patient had a lumbar fusion at l4-l5 using the matrix system on an unknown date approximately one year ago.Patient presented to surgeon complaining of postoperative pain on an unknown date.X-ray taken on an unknown date determined that a matrix top-loading polyaxial head had popped off one of the matrix screws.A revision surgery was scheduled for (b)(6) 2018, and the surgeon removed the entire construct on the right-hand side, and replaced it with another matrix construct.There was no allegation of deficiency against the hardware on the left side.There was no surgical delay, no fragments generated, and the patient outcome was reportedly good.There are two devices involved in this complaint.Concomitant devices reported: titanium matrix screw (part # unknown, lot # unknown, qty.1); titanium matrix top-loading polyaxial head, (part # 04.632.001, lot # unknown, qty.1); titanium matrix locking cap (part # 04.632.000, lot # unknown, qty.2); curved rod (part # unknown, lot # unknown, qty.1).This report is for one (1) ti matrix top loading polyaxial head.This is report 1 of 2 for complaint (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Customer quality conducted a review of x-rays provided.One polyaxial head (04.632.001) and one screw (part number unknown) were reported with the complaint that a matrix top-loading polyaxial head had popped off one of the matrix screws.However, the devices were not returned for investigation; only medical imaging was provided.Thus, the following investigation is based on the medical imaging.Device condition: the medical images appear to show a pair of rod segments each with two polyaxial screws and one rod.One of the polyaxial heads appears to have disassembled from the polyaxial screw.From the provided images, no fractures in the parts could be identified.Therefore, the overall complaint condition is confirmed and consistent with the reported condition as the polyaxial head appears to have popped off.Replication of this complaint condition is not applicable as the devices were not received for physical inspection.Lot number review: a dhr review could not be performed as the lot number is unknown.Drawing review: the following drawing was reviewed; ti top loading poly head, matrix: during the investigation no product design issues or discrepancies were observed that may have contributed to the complaint condition.As the physical device was not received no further dimensional or material testing could be completed.No definitive root cause was able to be determined as circumstances surrounding the event and over the implant duration are unknown and the devices were not available for inspection.Risk assessment review: the matrix 5.5 design and clinical risk management (dcrm) document was reviewed in action (b)(4) and found to adequately address the complaint condition.Conclusion: a visual inspection via the provided x-ray, drawing review, complaint history review, and risk assessment review were performed as part of this investigation.The complaint condition is confirmed as the provided images show the polyaxial head slightly disassembled from the screw.From the provided information, no design or manufacturing issue was identified.No new malfunctions were observed during the course of this investigation.The risk assessment was found to adequately address the complaint condition.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Search Alerts/Recalls
|
|
|