MEDTRONIC SOFAMOR DANEK KYPHON EXPRESS II INFLATABLE BONE TAMP; ARTHROSCOPE Back to Search Results
Catalog Number KE152
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 01/29/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Neither the device nor applicable imaging films were returned to manufacturer for evaluation therefore we are unable to determine the definitive cause of event.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
It was reported that the patient underwent kyphoplasty procedure at l4/l5 due to vertebral compression fractures.Intra-op, a 15mm inflatable bone tamp (ibt) ruptured on the left side of l5.As per the surgeon there must have been a bone shard or something hard within the left side of l5 that ruptured the ibt that was inserted.The procedure was sucessfully completed with the original product.No fragments of the ibt remained iniside the patient.No patient complications were reported as a result of the event.
 
Additional Manufacturer Narrative
Product analysis: the inflatable bone tamp (ibt) had been cut the full circumference of the balloon perpendicular to the shaft of the ibt.This type of damage is consistent with the balloon being cut by bone in the disc space then being damaged while retracted through the tube.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
KYPHON EXPRESS II INFLATABLE BONE TAMP
Common Device Name
ARTHROSCOPE
Manufacturer (Section D)
MEDTRONIC SOFAMOR DANEK
1800 pyramid place
memphis TN 38132
MDR Report Key7271051
Report Number1030489-2018-00222
Device Sequence Number2378270
Product Code HRX
UDI-Device Identifier00643169143777
UDI-Public00643169143777
Combination Product (Y/N)N
Initial Reporter StateMA
Initial Reporter CountryUS
PMA/510(K) Number
K123771
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 05/29/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberKE152
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/18/2018
Initial Date Received by Manufacturer 01/29/2018
Supplement Date Received by Manufacturer04/30/2018
Initial Report FDA Received Date02/14/2018
Supplement Report FDA Received Date05/29/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-