BARD PERIPHERAL VASCULAR, INC. DORADO PTA BALLOON DILATATION CATHETER Back to Search Results
Catalog Number DR4054
Medical Device Problem Codes Break (1069); Leak/Splash (1354)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 11/24/2017
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device history records were reviewed with special attention to the raw materials, subassemblies, manufacturing process and quality control testing.This lot met all release criteria.There was nothing found to indicate there was a manufacturing related cause for this event.This is the only event reported to date for this lot number and failure mode.Investigation summary: the device was returned for evaluation.An inflation attempt was made and a leak was noted, identifying a partial break at the butt joint between the balloon and catheter shaft.Therefore, the investigation was confirmed for the reported leak, as well as for a break at the butt joint.The break at the glue butt joint resulted in the identified leak.However, the definitive root cause for the partial break could not be determined based upon available information.It was unknown whether procedural issues contributed to the event.Labeling review: the current ifu (instructions for use) states: warnings: if resistance is met during manipulation, determine the cause of the resistance before proceeding.Applying excessive force to the catheter can result in tip breakage or balloon separation.
 
Event or Problem Description
It was reported that during an angioplasty procedure, the pta balloon catheter allegedly would not fully inflate during the first inflation.The health care provider removed the balloon catheter from the patient and made another inflation attempt outside of the body which allegedly identified a leak at the proximal catheter-balloon butt joint.There was no reported patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DORADO PTA BALLOON DILATATION CATHETER
Common Device Name
PTA BALLOON DILATATION CATHETER
Manufacturer (Section D)
BARD PERIPHERAL VASCULAR, INC.
1625 w 3rd st.
tempe AZ 85281
MDR Report Key7271232
Report Number2020394-2018-00096
Device Sequence Number1276188
Product Code DQY
UDI-Device Identifier00801741058486
UDI-Public(01)00801741058486
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K072283
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 02/14/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/21/2020
Device Catalogue NumberDR4054
Device Lot Number93UB0089
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/20/2017
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/24/2018
Initial Report FDA Received Date02/14/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-