• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TRIVIDIA HEALTH, INC. TRUETRACK; BLOOD GLUCOSE SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TRIVIDIA HEALTH, INC. TRUETRACK; BLOOD GLUCOSE SYSTEM Back to Search Results
Model Number TRUETRACK
Medical Device Problem Code High Test Results (2457)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 01/26/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Returned meter evaluated with no defect found.Test strips not returned for evaluation.Most likely underlying root cause: mlc-58-user had an inaccurate reference: self: the person is using themselves as the reference or how they feel at the time they run the blood test.Note: manufacturer contacted customer on (b)(6) 2018 in a follow-up call in order to ensure that the replacement products resolved the initial concern; customer stated that the replacement product is working to their satisfaction and they have not had any medical intervention since the last call.
 
Event or Problem Description
Consumer reported complaint for high blood glucose test results.The customer is concerned with tests results from results obtained of 501 and 232 mg/dl.The customer's expected fasting blood glucose test result range is 100 - 125 mg/dl.The customer feels well and did not report any symptoms.Medical attention is not reported as a result of the actual blood glucose results.During the call on (b)(6) 2018, a back to back blood test was performed by the customer fasting and produced test results of 196 mg/dl and 212 mg/dl using truetrack meter.The product is stored according to specification in the family room.Test strip lot manufacturer's expiration date is 05/31/2019 and open vial date is (b)(6) 2017.The meter memory was reviewed for previous test result history: result 1 :232mg/dl date: (b)(6) 2018time:09:02am fasting, result 2 :117mg/dl date: (b)(6) 2018time:02:48am fasting, result 3 :137mg/dl date: (b)(6) 2018time:02:40am fasting, result 4 :128mg/dl date: (b)(6) 2018time:02:48am fasting, result 5 :501mg/dl date: (b)(6) 2018time:08:24am fasting.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRUETRACK
Common Device Name
BLOOD GLUCOSE SYSTEM
Manufacturer (Section D)
TRIVIDIA HEALTH, INC.
2400 nw 55th court
fort lauderdale FL 33309
MDR Report Key7271596
Report Number1000113657-2018-00190
Device Sequence Number1801755
Product Code NBW
UDI-Device Identifier00021292008888
UDI-Public(01)00021292008888(17)190531(10)RT5054
Combination Product (Y/N)N
Initial Reporter StateCA
PMA/510(K) Number
K032657
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer
Initial Reporter Occupation Other
Remedial Action Replace
Type of Report Initial
Report Date (Section B) 02/14/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date05/31/2019
Device Model NumberTRUETRACK
Device Catalogue NumberA4H01-81
Device Lot NumberRT5054
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/08/2018
Was the Report Sent to FDA? No
Event Location Home
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/26/2018
Initial Report FDA Received Date02/14/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-