Model Number MMT-723LNAP
Health Effect - Clinical Code Hyperglycemia (1905)
Date of Event 02/02/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
 
Event or Problem Description
Customer reported via phone call for blank display and cosmetic damage.Blood glucose at the time of incident was unknown.Customer reported blank display and mentioned that insulin pump was not working.The customer reported cracked pieces around the battery compartment.The insulin pump will not be returned for analysis.
 
Additional Manufacturer Narrative
Pump had blank display due to corrosion on the interface board.Unable to perform idle current test, run current test, self test, off no power test, basic occlusion test, occlusion test, prime test, excessive no delivery test and displacement test due to blank display.Pump had broken belt clip slot.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP MMT-723LNAP
Common Device Name
PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR
Manufacturer (Section D)
MEDTRONIC MINIMED
18000 devonshire st.
northridge CA 91325 1219
MDR Report Key7271834
Report Number2032227-2018-01491
Device Sequence Number10796928
Product Code OYC
UDI-Device Identifier00613994904492
UDI-Public(01)00613994904492
Combination Product (Y/N)N
Initial Reporter StateIN
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer
Initial Reporter Occupation Patient
Type of Report Initial,Followup
Report Date (Section B) 08/10/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Model NumberMMT-723LNAP
Device Catalogue NumberMMT-723LNAP
Device Lot NumberA000202976
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/27/2018
Initial Date Received by Manufacturer 02/02/2018
Supplement Date Received by Manufacturer07/13/2018
Initial Report FDA Received Date02/14/2018
Supplement Report FDA Received Date08/10/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age38 YR
Patient Weight220
-
-