ABBOTT MEDICAL OPTICS TECNIS SYMFONY TORIC; MULTIFOCAL IOLS Back to Search Results
Model Number ZXT150
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Retinal Detachment (2047); Vitrectomy (2643); No Code Available (3191)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Date of event: unknown, not provided, but the best estimate date is (b)(6) 2017, as retina detachment was reported as 6 months post op.(b)(4).All pertinent information available to abbott medical optics has been submitted.
 
Event or Problem Description
It was reported that tecnis symfony toric intraocular lens (model zxt150, +17.5 diopter) was explanted in a secondary procedure.Reportedly, there was retinal detachment 6 months post op, which required a secondary procedure.Lens was explanted and a vitrectomy performed.An anterior chamber intraocular lens was used as replacement lens for patient.Patient is in stable condition.No further information provided.
 
Additional Manufacturer Narrative
Device available for evaluation ¿ yes, returned to manufacturer on 3/21/2018.Device returned to manufacturer ¿ yes.Device evaluation: the intraocular lens (iol) was returned to the manufacturing site for evaluation.The sample was inspected by a qualified final inspection operator using 12x magnification.It can be seen that there is a scratch on the optic, most probably to make explant possible.The customer's reported complaint could not be verified.Manufacturing records review: the manufacturing records for the intraocular lens were reviewed.The product was manufactured and released according to specification.A search on complaints revealed that no similar complaints were received for this production order number.Labeling review: the directions for use (dfu) was reviewed.The dfu adequately provides instructions and precautions along with warnings for the proper use and handling of the device.As a result of the investigation there is no indication of a product quality deficiency and the reported issue could not be verified.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TECNIS SYMFONY TORIC
Common Device Name
MULTIFOCAL IOLS
Manufacturer (Section D)
ABBOTT MEDICAL OPTICS
santa ana CA
MDR Report Key7272043
Report Number9614546-2018-00149
Device Sequence Number1847414
Product Code POE
UDI-Device Identifier05050474605152
UDI-Public(01)05050474605152(17)211007
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
P980040
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Device Explanted Year2017
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 06/21/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Physician
Device Expiration Date10/07/2021
Device Model NumberZXT150
Device Catalogue NumberZXT150U175
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/21/2018
Initial Date Received by Manufacturer 01/18/2018
Supplement Date Received by Manufacturer06/04/2018
Initial Report FDA Received Date02/14/2018
Supplement Report FDA Received Date06/21/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-