Model Number U420048RX
Medical Device Problem Codes Material Puncture/Hole (1504); Material Rupture (1546)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 12/21/2017
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
No hospital/medical records or medical images have been made available to the manufacturer.As the lot number for the device was provided, a review of the device history records is currently being performed.The device has been returned to the manufacturer for evaluation.The investigation of the reported event is currently underway.The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.
 
Event or Problem Description
It was reported that during an angioplasty procedure of a vein graft in the left posterior tibial, the pta balloon allegedly ruptured at 8atm.There was no reported retraction issues.The pta balloon was exchanged over the guidewire for another that was used to complete the procedure.There was no reported patient injury.
 
Additional Manufacturer Narrative
Manufacturing review: the device history records were reviewed with special attention to the raw materials, subassemblies, manufacturing process and quality control testing.This lot met all release criteria.There was nothing found to indicate there was a manufacturing related cause for this event.This is the only complaint reported to date for this lot number and failure mode.Investigation summary: the investigation is inconclusive for the reported rupture, as the device was not returned for evaluation.The definitive root cause for the reported rupture could not be determined based upon available information.It is unknown whether patient and/or procedural issues contributed to the event.Labeling review: the current ifu (instructions for use) states: warnings: when the catheter is exposed to the vascular system, it should be manipulated while under high quality fluoroscopic observation.Do not advance or retract the catheter unless the balloon is fully deflated.If resistance is met during manipulation, determine the cause of the resistance before proceeding.Applying excessive force to the catheter can result in tip or catheter breakage, catheter kink, or balloon separation.If resistance is felt during post procedure withdrawal of the catheter through the introducer sheath/guide catheter, determine if contrast is trapped in the balloon with fluoroscopy.If contrast is present, push the balloon out of the introducer sheath/guide catheter and then completely evacuate the contrast before proceeding to withdraw the balloon.If resistance is still felt during post procedure withdrawal of the catheter, it is recommended to remove the balloon catheter and introducer sheath/guide catheter as a single unit.Do not exceed the rated burst pressure.To prevent over pressurization, use of a pressure monitoring device is recommended.Balloon rupture may occur if the rbp rating is exceeded.
 
Event or Problem Description
It was reported that during an angioplasty procedure of a vein graft in the left posterior tibial, the pta balloon allegedly ruptured at 8atm.There was no reported retraction issues.The pta balloon was exchanged over the guidewire for another that was used to complete the procedure.There was no reported patient injury.
 
Additional Manufacturer Narrative
After further review of the additional event details provided, the event suggests an uncomplicated balloon rupture.This event was reassessed and determined to be not mdr reportable.However, an initial mdr has already been submitted; therefore, the purpose of this supplemental report is to document the change in reportability classification.
 
Event or Problem Description
It was reported that during an angioplasty procedure of a vein graft in the left posterior tibial, the pta balloon allegedly ruptured at 8atm.There was no reported retraction issues, no detachment of balloon material, and no vessel damage.The pta balloon was exchanged over the guidewire for another that was used to complete the procedure.There was no reported patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ULTRAVERSE RX PTA DILATATION CATHETER
Common Device Name
PTA BALLOON DILATATION CATHETER
Manufacturer (Section D)
BARD PERIPHERAL VASCULAR, INC.
1625 w 3rd st.
tempe AZ 85281
MDR Report Key7272964
Report Number2020394-2018-00106
Device Sequence Number1028239
Product Code LIT
UDI-Device Identifier00801741057533
UDI-Public(01)00801741057533
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
PMA/510(K) Number
K131199
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,user f
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 03/16/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/30/2020
Device Model NumberU420048RX
Device Catalogue NumberU420048RX
Device Lot NumberCMBU0366
Was Device Available for Evaluation? No
Date Returned to Manufacturer01/04/2018
Initial Date Received by Manufacturer 01/19/2018
Supplement Date Received by Manufacturer03/09/2018
03/09/2018
Initial Report FDA Received Date02/15/2018
Supplement Report FDA Received Date03/14/2018
03/16/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age63 YR
Patient Weight142
-
-