| Catalog Number |
150432 |
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.The device was not returned.
|
| |
|
Event or Problem Description
|
|
It was reported that the urine would not drain from the catheter into the leg bag unless it was a brand new bag.The complainant alleged that he disconnected the extension tubing from the leg bag to allow air into the system to help urine flow, but urine still would not drain into the bag.The complainant only experienced this problem after washing the leg bag and trying to reuse it.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was not returned for evaluation.The device history record was reviewed and found nothing that could have caused or contributed to the reported event.The instructions for use state the following: ¿directions for use: 1.Separate notches within circles.Pull straps through holes and around leg.2.Position bag on leg with flutter valve at top.3.Attach catheter or extension tubing to top inlet.When wearing bag below knee, attach bard extension tubing (catalog no.150615 or 4a4194).4.To empty dispoz-a-bag, push green lever on flip-flo valve out and down.Important: be sure to reclose flip-flo valve after emptying bag.5.After use, this product may be a potential biohazard.Handle and dispose of in accordance with accepted medical practice and applicable, local, state and federal laws and regulations." (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that the urine would not drain from the catheter into the leg bag unless it was a brand new bag.The complainant alleged that he disconnected the extension tubing from the leg bag to allow air into the system to help urine flow, however urine still would not drain into the bag.The complainant only experienced this problem after washing the leg bag and trying to reuse it.
|
| |
|