Device Problems Signal Artifact/Noise (1036); Degraded (1153)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
Note: we have not completed our investigation of this event.We will file a follow-up emdr at the completion of the investigation.Internal cross reference: complaint pr# (b)(4).
 
Event Description
Engineering analysis concluded that this event has the potential to result in image misinterpretation due to an artifact from a degraded compensator within the collimator.Therefore, this issue has been determined to be a reportable event.
 
Manufacturer Narrative
The customer reported during system constancy test, eight images display and image 1 and image 5 show a hounsfield unit (hu) value of 23-24.Based on further investigation and evaluation of the system by a philips field service engineer (fse), it was determined that an issue with the system collimator was not the cause of the image quality as originally reported.The cause of the hu value issue was caused by a corrupt air calibration file stored in the reconstruction calibration tables.The file was reloaded by the fse and the constancy tests passed.The hu values were now within specification.There was no report of patient impact.The system is in clinical use.This issue was determined not to be a reportable event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BRILLIANCE AIR 40/64/UCT
Type of Device
SYSTEM, X-RAY, TOMOGRAPHY, COMPUTED
Manufacturer (Section D)
PHILIPS MEDICAL SYSTEMS (CLEVELAND), INC.
595 miner rd
cleveland OH 44143
MDR Report Key7273408
MDR Text Key100441639
Report Number1525965-2018-00351
Device Sequence Number1
Product Code JAK
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 11/16/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/15/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Other Device ID NumberN/A
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received11/16/2016
Is This a Reprocessed and Reused Single-Use Device? No
Removal/Correction NumberN/A
Patient Sequence Number1
-
-