PHILIPS MEDICAL SYSTEMS (CLEVELAND), INC. BRILLIANCE AIR 40/64/UCT; SYSTEM, X-RAY, TOMOGRAPHY, COMPUTED
|
Back to Search Results |
|
Device Problems
Signal Artifact/Noise (1036); Degraded (1153)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Type
malfunction
|
Manufacturer Narrative
|
Note: we have not completed our investigation of this event.We will file a follow-up emdr at the completion of the investigation.Internal cross reference: complaint pr# (b)(4).
|
|
Event Description
|
Engineering analysis concluded that this event has the potential to result in image misinterpretation due to an artifact from a degraded compensator within the collimator.Therefore, this issue has been determined to be a reportable event.
|
|
Manufacturer Narrative
|
The customer reported during system constancy test, eight images display and image 1 and image 5 show a hounsfield unit (hu) value of 23-24.Based on further investigation and evaluation of the system by a philips field service engineer (fse), it was determined that an issue with the system collimator was not the cause of the image quality as originally reported.The cause of the hu value issue was caused by a corrupt air calibration file stored in the reconstruction calibration tables.The file was reloaded by the fse and the constancy tests passed.The hu values were now within specification.There was no report of patient impact.The system is in clinical use.This issue was determined not to be a reportable event.
|
|