The customer experienced an issue over several days where the elecsys free psa immunoassay would be ordered in the cobas e 411 immunoassay analyzer software for patient samples and qc material where the test had not been requested and the result for free psa was identical for all patient samples.The field service representative could not find a cause.Operational and mechanical checks were acceptable.The customer powered down and rebooted the analyzer multiple times.The customer's laboratory information system (lis) was also restarted.The instrument was then running as expected and the problem seemed to be resolved.On (b)(6) 2018, the customer repeated all of the patient samples.Two examples were provided.Patient 1 original result was 39.98 ng/ml and the repeat result was 1.87 ng/ml.Patient 2 original result was 39.98 ng/ml and the repeat result was 0.488 ng/ml.No erroneous results were reported outside of the laboratory.The repeat results were believed to be correct.No patients were adversely affected.The free psa reagent lot number was 243142.The expiration date was requested but was not provided.The investigation confirmed the software issue which has been corrected in a new software version.
|