Model Number PM3262
Device Problem Device Difficult to Program or Calibrate (1496)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 01/17/2018
Event Type  Injury  
Event Description
Following the initial implant of the device, the device was unable to be programmed.The patient¿s hemodynamic state worsened as a result and the device was explanted and replaced.The patient was stable following the procedure.
 
Manufacturer Narrative
Interrogation of the device throughout lab testing revealed no anomalies.Multiple telemetry tests were performed and were successful.Further testing found the device exhibited normal characteristics.A longevity assessment was performed and device was in the normal range of operation with appropriate remaining longevity.The reported event of inability to program could not be confirmed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUADRA ALLURE MP RF CRT-P
Type of Device
IMPLANTABLE PACEMAKER PULSE GENERATOR
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.
15900 valley view court
sylmar CA 91342
Manufacturer Contact
robert greenleaf
15900 valley view court
sylmar, CA 91342
8184932577
MDR Report Key7273936
MDR Text Key100179814
Report Number2017865-2018-02488
Device Sequence Number1
Product Code NKE
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
P910023
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 05/23/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/15/2018
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date04/30/2019
Device Model NumberPM3262
Device Lot Number6256395
Other Device ID Number05414734509091
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/17/2018
Is the Reporter a Health Professional? Yes
Date Manufacturer Received05/14/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/17/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
-
-