Catalog Number 375154
Medical Device Problem Code Fluid/Blood Leak (1250)
Health Effect - Clinical Code No Patient Involvement (2645)
Date of Event 01/20/2016
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Event 2: (b)(4).Two (2) used samples were received in conjunction with this complaint.The samples were visually examined and no manufacturing defects were noted.The samples were physically tested, for leakage, with failing results.Leakage was observed at the bonded joint of the tubing and the blood filter housing assembly.Therefore, the reported defect was confirmed.Review of the discrepancy management system database performed for the reported lot number did not reveal any abnormalities or non-conformances of this nature.To address this issue, a modified fixture design was implemented which increased the fixture's diameter and pin to allow for easier application of the solvent.We will maintain this report for further references and continue to monitor other reports for similar occurrences.Note: this case is being filed retrospectively as a result of a review of recent customer complaint information.Based on a new complaint case and an updated procedure, it was determined that this case is reportable in accordance with the requirements of 21 cfr 803.B.Braun has conducted a retrospective review for all complaints of similar nature in accordance with b.Braun internal procedures.
 
Event or Problem Description
As reported by the user facility: event 2: during priming, the nurse observed leaking of lactaced ringers at the lower y-port.The nurse was able to start administration with another blood set.No delay in therapy.No injuries reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BLOOD ADMINISTRATION SET
Common Device Name
SET, ADMINISTRATION, INTRA
Manufacturer (Section D)
B. BRAUN DOMINICAN REPUBLIC INC.
las americas industrial park
km22 autopista las americas
santo domingo,
DR 
MDR Report Key7274183
Report Number9614279-2018-00017
Device Sequence Number123526
Product Code FPA
UDI-Device Identifier04046964182990
UDI-Public04046964182990
Combination Product (Y/N)N
Initial Reporter StateGA
Initial Reporter CountryUS
PMA/510(K) Number
K955585
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional,user faci
Initial Reporter Occupation Health Professional
Type of Report Initial
Report Date (Section B) 02/15/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/30/2020
Device Catalogue Number375154
Device Lot Number0061453625
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/27/2016
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/21/2016
Initial Report FDA Received Date02/15/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-