B. BRAUN DOMINICAN REPUBLIC INC. BLOOD ADMINISTRATION SET; SET, ADMINISTRATION, INTRA Back to Search Results
Catalog Number 490453
Medical Device Problem Code Fluid/Blood Leak (1250)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This report has been identified as (b)(4).One (1) sample, in open packaging, was returned for evaluation.Visual examination noted the ball lodged between the tubing dispensed and roam freely in the tubing.A leakage test was performed per specification and leakage was found from the blood filter.All information of this incident was forwarded to our supplier for evaluation.Our supplier performed an investigation, but could not determine the exact root cause into the incident.However, since the production of the lot, they have added a new electronic solvent dispenser to assure better solvent application to the bonded area.In addition, the supplier has also increased their in-process inspection sample size to improve detection of defects.Review of the discrepancy management system database performed for the reported lot number did not reveal any abnormalities or non-conformances of this nature.We will maintain this report for further references and continue to monitor other reports for similar occurrences.Note: this case is being filed retrospectively as a result of a review of recent customer complaint information.Based on a new complaint case and an updated procedure, it was determined that this case is reportable in accordance with the requirements of 21 cfr 803.B.Braun has conducted a retrospective review for all complaints of similar nature in accordance with b.Braun internal procedures.
 
Event or Problem Description
As reported by the user facility: end user reports that the iv tube cracked just below the filter aspect of the tubing, when clinician was using hand pump.Sodium chloride was used at the time of incident.No injuries to patient reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BLOOD ADMINISTRATION SET
Common Device Name
SET, ADMINISTRATION, INTRA
Manufacturer (Section D)
B. BRAUN DOMINICAN REPUBLIC INC.
las americas industrial park
km22 autopista las americas
santo domingo,
DR 
MDR Report Key7274446
Report Number9614279-2018-00015
Device Sequence Number1177437
Product Code FPA
UDI-Device Identifier04046964848957
UDI-Public04046964848957
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
PMA/510(K) Number
K083723
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Health Professional
Type of Report Initial
Report Date (Section B) 02/15/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/31/2019
Device Catalogue Number490453
Device Lot Number0061517190
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/21/2016
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/09/2016
Initial Report FDA Received Date02/15/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Congenital Anomaly;
-
-