MAUDE Adverse Event Report: BARD PERIPHERAL VASCULAR, INC. FLUENCY PLUS ENDOVASCULAR STENT GRAFT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BARD PERIPHERAL VASCULAR, INC. FLUENCY PLUS ENDOVASCULAR STENT GRAFT Back to Search Results
Model Number FEM14120
Medical Device Problem Codes Positioning Failure (1158); Fracture (1260); Misfire (2532)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 01/18/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
No medical records or no medical images have been made available to the manufacturer; however a photo was provided and a photo review will be performed.As the lot number for the device was provided, a review of the device history records is currently being performed.The device has been returned to the manufacturer for evaluation.The investigation of the reported event is currently underway the information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.
 
Event or Problem Description
It was reported that during a stent graft placement procedure for an occlusion in the iliac artery, the outer sheath of the catheter allegedly fractured and the stent graft allegedly partially deployed.Reportedly, the device and the delivery system were removed without incident and the procedure was concluded without any device implanted.There was no reported patient injury.
 
Event or Problem Description
It was reported that during a stent graft placement procedure for an occlusion in the iliac artery, the outer sheath of the catheter allegedly fractured and the stent graft allegedly partially deployed.Reportedly, the device and the delivery system were removed without incident and the procedure was concluded without any device implanted.There was no reported patient injury.
 
Additional Manufacturer Narrative
Manufacturing review: a manufacturing review was performed.The lot records have been reviewed with special attention to the manufacturing and inspection of this product and the product was found to have met the specification prior to shipment.The review shows that no remarkable incidents occurred during the manufacturing process.No relevant manufacturing process changes were implemented that could have led to the event reported.Investigation summary: based on the condition of the returned sample, the reported issue could be confirmed.The stent graft was found to be partially deployed and the outer sheath was found to be elongated indicating that increased friction must have affected on the delivery system during the attempt to deploy the stent graft finally leading to the fracture of the outer sheath which made a stent graft deployment impossible.Potential factors which may have caused or contributed to the reported issue have been considered.Based on the information available and the evaluation of the sample returned, a definite root cause for the reported event could not be determined.Labeling review: in reviewing the labeling supplied with this product it was found that the instructions for use (ifu) sufficiently address the potential risks.Regarding preparation of the device the ifu states: "prior to loading the endovascular system over a guide wire, both ports must be flushed with sterile saline (.).Flushing these lumens will also facilitate stent graft deployment." regarding the anatomy of the placement site the ifu states: "the safety and effectiveness of the device when placed across a tight bend (.) has not been evaluated.Prior to stent graft deployment, ensure that the proximal (inflow) stent graft end is positioned in a straight section of the lumen to reduce the risk of higher deployment forces and possible endovascular system failure".Furthermore, the ifu states: "if unusual resistance or high deployment force is encountered during stent graft deployment, abort the procedure, remove the delivery system and use an alternative device.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FLUENCY PLUS ENDOVASCULAR STENT GRAFT
Common Device Name
ENDOVASCULAR STENT GRAFT
Manufacturer (Section D)
BARD PERIPHERAL VASCULAR, INC.
1625 w 3rd st.
tempe AZ 85281
MDR Report Key7275122
Report Number2020394-2018-00115
Device Sequence Number164430
Product Code PFV
UDI-Device Identifier04049519008721
UDI-Public(01)04049519008721
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
PMA/510(K) Number
P130029
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 03/14/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/24/2019
Device Model NumberFEM14120
Device Catalogue NumberFEM14120
Device Lot NumberANAQ3433
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/14/2018
Initial Date Received by Manufacturer 01/23/2018
Supplement Date Received by Manufacturer03/08/2018
Initial Report FDA Received Date02/15/2018
Supplement Report FDA Received Date03/14/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age43 YR
Patient Weight59
-
-