EDWARDS LIFESCIENCES CARPENTIER-EDWARDS PHYSIO TRICUSPID ANNULOPLASTY RING; RING, ANNULOPLASTY Back to Search Results
Model Number 6200
Device Problem Appropriate Term/Code Not Available (3191)
Patient Problem Tricuspid Regurgitation (2112)
Event Date 01/09/2018
Event Type  Injury  
Manufacturer Narrative
(b)(4).Ring dehiscence may occur early or late.When it occurs in the early post-operative period, it is typically a result of an inadequate valve repair in combination with friable myocardial tissue.Late dehiscence can occur as a result of successive dilatation of cardiac structures that result from progression of disease or from endocarditis.In this case, the ring was downsized two sizes, suggesting that this event was due to mis-sizing at the initial time of implantation.The root cause of this event cannot be conclusively determined; however, it appears that this event was due to procedural related factors.The device history record (dhr) was not reviewed as the reported event does not allege a malfunction that could be related to a manufacturing deficiency and/or one was not confirmed through investigation.Edwards lifesciences will continue to monitor all reported events.No further actions are required at this time.
 
Event Description
Edwards lifesciences maintains an implant patient registry.This registry is a patient tracking mechanism for serialized edwards implantable devices (bioprosthetic heart valves and annuloplasty rings), rather than a true post-market surveillance registry.Through the registry, edwards is notified when these devices are implanted.In addition, patient and/or device status may be reported to the registry via the implantation data cards.The information is received from various sources (e.G.Surgeon, hospital and patient family members) and is not received in the form of a conventional "customer complaint." the information reported may or may not be related to the edwards device.In this case, it was learned that a 30 mm tricuspid annuloplasty ring, implanted two (2) years and six (6) months, was explanted due to tricuspid regurgitation and partial dehiscence of the ring.Per medical records, this case involved a (b)(6)-year-old female with a history of aortic, mitral and tricuspid valve repair.She presented with severe recurrent tricuspid regurgitation and severe aortic insufficiency.Echocardiogram confirmed partial dehiscence of the tricuspid ring with severe tricuspid regurgitation.Intraoperatively, there was evidence of partial tricuspid ring dehiscence.The ring was excised and a smaller 26 mm tricuspid ring was seated and sutured into place with corknot device.Aortic valve was also performed.She was weaned from cpb.She was transferred to icu in satisfactory condition.The patient was discharged to a nursing facility on pod #10 in stable condition.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARPENTIER-EDWARDS PHYSIO TRICUSPID ANNULOPLASTY RING
Type of Device
RING, ANNULOPLASTY
Manufacturer (Section D)
EDWARDS LIFESCIENCES
1 edwards way
irvine CA 92614
Manufacturer (Section G)
EDWARDS LIFESCIENCES LLC
1 edwards way
irvine CA 92614
Manufacturer Contact
neil landry
1 edwards way
mle-2
irvine, CA 92614
9492502289
MDR Report Key7275298
MDR Text Key100231423
Report Number2015691-2018-00540
Device Sequence Number1
Product Code KRH
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K103520
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,other
Reporter Occupation Physician
Type of Report Initial
Report Date 01/26/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/15/2018
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date02/19/2019
Device Model Number6200
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/26/2018
Date Device Manufactured02/25/2014
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
Patient Age62 YR
-
-