• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC IRELAND RESOLUTE INTEGRITY RX; STENT, CORONARY, DRUG-ELUTING

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC IRELAND RESOLUTE INTEGRITY RX; STENT, CORONARY, DRUG-ELUTING Back to Search Results
Catalog Number RSINT30018UX
Medical Device Problem Code Material Deformation (2976)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 01/19/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
It was reported that a resolute integrity drug eluting stent was frayed.This occurred in vivo after balloon angioplasty.No injury to the patient was reported.
 
Additional Manufacturer Narrative
Evaluation summary: the stent was positioned on the balloon between the marker bands as per specifications.Misalignment was evident to the 1st stent wrap, with strut raised.Deformation was evident to the distal tip, with tip appearing torn.If information is provided in the future, a supplemental report will be issued.
 
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RESOLUTE INTEGRITY RX
Common Device Name
STENT, CORONARY, DRUG-ELUTING
Manufacturer (Section D)
MEDTRONIC IRELAND
parkmore business park west
galway
MDR Report Key7276039
Report Number9612164-2018-00363
Device Sequence Number19151448
Product Code NIQ
UDI-Device Identifier00643169037151
UDI-Public00643169037151
Combination Product (Y/N)N
Initial Reporter StateAZ
Initial Reporter CountryUS
PMA/510(K) Number
P110013
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup,Followup
Report Date (Section B) 10/05/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/31/2019
Device Catalogue NumberRSINT30018UX
Device Lot Number0008657150
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/13/2018
Initial Date Received by Manufacturer 01/22/2018
Supplement Date Received by Manufacturer04/03/2018
06/25/2018
08/08/2018
Initial Report FDA Received Date02/16/2018
Supplement Report FDA Received Date05/01/2018
07/23/2018
10/05/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-