MEDTRONIC IRELAND RESOLUTE INTEGRITY RX; STENT, CORONARY, DRUG-ELUTING Back to Search Results
Catalog Number RSINT22530UX
Medical Device Problem Code Device Dislodged or Dislocated (2923)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 01/23/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
A resolute integrity rx drug eluting stent was attempted to be used in a procedure.There was no damage noted to packaging and no issues when removing the device from the hoop.The device was inspected with no issues.The lesion was pre-dilated.During the procedure, the device did not pass through a previously deployed stent.Resistance was encountered while advancing the device.A guideliner was used but it still did not aid delivery.While being removed from the patient, the stent was pulled back into the guideliner and the stent was reported to have fell off the delivery system balloon.The dislodged stent was removed.No patient injury was reported.
 
Additional Manufacturer Narrative
Evaluation summary: the stent was not present on the balloon and was returned for analysis.Deformation was evident to the entire distal stent wrap with struts raised, bunched and overlapping.Crimp impressions were visible on the exposed balloon surface.The balloon folds remained intact.No deformation was evident to the distal tip during analysis.If information is provided in the future, a supplemental report will be issued.
 
Additional Manufacturer Narrative
Additional information: no difficulties were noted when removing the protective sheath prior to use.The stent was inspected post removal of the protective sheath, with no issues noted.The inflation device remained on neutral pressure during delivery of the device.The stent was removed after being dislodged from the undeployed balloon.This occurred because stent slipped off balloon when being pulled back in non-medtronic catheter.The stent was caught on the edge of the non-medtronic catheter when being pulled back in.The.014¿ wire was still in the vessel.The entire system was pulled with the wire, non-medtronic catheter, and guide catheter along with stent out all at once if information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RESOLUTE INTEGRITY RX
Common Device Name
STENT, CORONARY, DRUG-ELUTING
Manufacturer (Section D)
MEDTRONIC IRELAND
parkmore business park west
galway
MDR Report Key7276102
Report Number9612164-2018-00366
Device Sequence Number1343418
Product Code NIQ
UDI-Device Identifier00643169037335
UDI-Public00643169037335
Combination Product (Y/N)N
Initial Reporter StateMI
Initial Reporter CountryUS
PMA/510(K) Number
P110013
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 05/08/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/08/2019
Device Catalogue NumberRSINT22530UX
Device Lot Number0008669537
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/07/2018
Initial Date Received by Manufacturer 01/23/2018
Supplement Date Received by Manufacturer02/23/2018
04/09/2018
Initial Report FDA Received Date02/16/2018
Supplement Report FDA Received Date03/22/2018
05/08/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-