• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC MINIMED PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP MMT-723RNAP; PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC MINIMED PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP MMT-723RNAP; PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR Back to Search Results
Model Number MMT-723RNAP
Health Effect - Clinical Code Hyperglycemia (1905)
Date of Event 02/05/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
 
Event or Problem Description
The customer reported via phone call the insulin pump alarmed motor error.The customer¿s blood glucose level was 308 mg/dl at the time of the incident and treated with manual injection.Customer reported that the insulin pump alarmed motor error after possible exposure to mri.The customer stated that the drive support cap was normal and that the insulin pump could have been exposed to high magnetic fields.Customer reported a motor position encoder error.The customer was advised to discontinue use of the insulin pump and to revert to the back-up plan.The insulin pump is being replaced and is expected to return for analysis.
 
Additional Manufacturer Narrative
Pump passed the rewind, basic occlusion, prime and displacement test.Pump was received with stuck motor error alarm loop during bolus delivery.The motor was tested outside of the device and passed.The motor may have had an intermittent failure that was not detected during our testing.Unable to confirm the motor position encoder error alarm due to erased history file.Pump had cracked case at display window corners, cracked battery tube threads, cracked reservoir tube lip, cracked reservoir tube window, cracked case at reservoir tube window corner, belt clip slot, minor scratched display window, stained end cap sticker and stained address/serial number label.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP MMT-723RNAP
Common Device Name
PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR
Manufacturer (Section D)
MEDTRONIC MINIMED
18000 devonshire st.
northridge CA 91325 1219
MDR Report Key7276720
Report Number2032227-2018-01522
Device Sequence Number1847579
Product Code OYC
UDI-Device Identifier00613994904393
UDI-Public(01)00613994904393
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer
Initial Reporter Occupation Patient
Type of Report Initial,Followup
Report Date (Section B) 07/16/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model NumberMMT-723RNAP
Device Catalogue NumberMMT-723RNAP
Device Lot NumberA000204715
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/19/2018
Initial Date Received by Manufacturer 02/05/2018
Supplement Date Received by Manufacturer06/20/2018
Initial Report FDA Received Date02/16/2018
Supplement Report FDA Received Date07/16/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age71 YR
Patient Weight253
-
-