During use, it was reported that the maxi ld ruptured.There was no patient injury.The target lesion is the right femoral vein which was slightly tortuous and calcified.The device was replaced with a non cordis balloon and the procedure was completed.The device was prep normally.There was no difficulty inflating the balloon during prep.There were no difficulties removing the product from the hoop, the protective balloon cover, the stylet or any of the sterile packaging components.There were no kinks or other damages noted prior to inserting the product the product into the patient.There was no damages noted after the device was removed from the patient.The device did not kinked anytime during the procedure.The contrast to saline ratio was "normal".Boston scientific inflation device was used, which was successfully used with other devices.There was no resistance/friction while inserting the balloon through the rotating hemostatic valve or through the guide catheter.The device was prepped according to the instruction for use (ifu).The product was removed intact (in one piece) from the patient.There was no damages noted to the device packaging.The device is not available for analysis as it was thrown away.No other information was provided.
|
|
During use, it was reported that the maxi ld ruptured.There was no patient injury.The target lesion is the right femoral vein which was slightly tortuous and calcified.The device was replaced with a non cordis balloon and the procedure was completed.The device was prep normally.There was no difficulty inflating the balloon during prep.There were no difficulties removing the product from the hoop, the protective balloon cover, the stylet or any of the sterile packaging components.There were no kinks or other damages noted prior to inserting the product the product into the patient.There were no damages noted after the device was removed from the patient.The device did not kink anytime during the procedure.The contrast to saline ratio was "normal".Boston scientific inflation device was used, which was successfully used with other devices.There was no resistance/friction while inserting the balloon through the rotating hemostatic valve or through the guide catheter.The device was prepped per the instruction for use (ifu).The product was removed intact (in one piece) from the patient.There were no damages noted to the device packaging.The device is not available for analysis as it was thrown away.No other information was provided.A device history record (dhr) review of lot 17642718 revealed no anomalies or non-conformances during the manufacturing and inspection processes that can be associated with the reported event.The reported ¿balloon burst-at/below rbp (peripheral)¿ could not be confirmed as the device was not returned for analysis.The exact cause could not be determined.Based on the limited information available for review, vessel characteristics of slightly tortuous and calcified may have contributed to the reported event.Per the instructions for use (ifu), which is not intended as a mitigation, per the instructions for use ¿prior to angioplasty, the catheter should be examined to verify functionality and ensure that its size and shape are suitable for the specific procedure for which it is to be used.Carefully advance the catheter through a sheath or through the percutaneous entry site.Note: gentle counterclockwise rotation of the balloon may ease introduction through the sheath or percutaneous entry site.Note: perform all further catheter manipulations under fluoroscopy.Carefully advance the catheter to the selected site.Caution: if strong resistance is met during advancement or withdrawal of the catheter, discontinue movement and determine the cause of resistance before proceeding.If the cause of resistance cannot be determined, withdraw the entire system.¿ neither the dhr nor the information available suggests a design or manufacturing related cause for the reported event; therefore, no corrective/preventive action will be taken.
|