Catalog Number 7N8377
Medical Device Problem Code Fitting Problem (2183)
Health Effect - Clinical Code No Patient Involvement (2645)
Date of Event 10/03/2017
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Should additional relevant information become available, a supplemental report will be submitted.
 
Event or Problem Description
It was reported that a one-link set did not properly connect to a catheter.This occurred before use.There was no patient involvement.No additional information is available.
 
Additional Manufacturer Narrative
One device was received for evaluation.During visual inspection, it was noted no extension was returned.Visual inspection did not identify any abnormalities that could have contributed to the reported condition.Clear passage, pressure and functional testing were all completed and the device performed according to product specifications.The reported condition was not verified.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EXTENSION SETS WITH ONE-LINK NEEDLE-FREE LV CONNECTOR
Common Device Name
SET, ADMINISTRATION, INTRAVASCULAR
Manufacturer (Section D)
BAXTER HEALTHCARE - DOMINICAN REPUBLIC
haina san cristobal
MDR Report Key7277058
Report Number1416980-2018-00817
Device Sequence Number1276210
Product Code FPA
UDI-Device Identifier00085412091808
UDI-Public(01)00085412091808
Combination Product (Y/N)N
Initial Reporter CountryCO
PMA/510(K) Number
K132734
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Pharmacist
Type of Report Initial,Followup
Report Date (Section B) 03/27/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Expiration Date07/12/2022
Device Catalogue Number7N8377
Device Lot NumberDR17F24024
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/22/2018
Initial Date Received by Manufacturer 01/24/2018
Supplement Date Received by Manufacturer03/02/2018
Initial Report FDA Received Date02/16/2018
Supplement Report FDA Received Date03/27/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-