| Model Number |
182263CS-V |
| Medical Device Problem Codes |
Detachment Of Device Component (1104); Difficult To Position (1467)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
01/14/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The lot number was provided.A review of the approved device history records indicated the lot met all release criteria.A lot history trending review was performed and there were no similar complaints for this lot number.The device was not returned for evaluation; therefore, a product analysis could not be performed.The root cause cannot be determined.
|
| |
|
Event or Problem Description
|
|
It was reported that it was difficult to place the embolization coil in the aneurysm.During removal, the coil detached in the microcatheter.The coil was removed in its entirety with the microcatheter.There was no reported intervention or patient injury.The patient is reported to be stable.
|
| |
|