The zoll catheter was not returned for investigation.Therefore, a physical investigation could not be performed.A supplemental report will be filed if the product is returned and investigation has been completed.Medical safety assessments: the use of endovascular cooling catheters was associated with a higher risk of venous catheter-related thrombosis compared with standard central venous catheters, with potential severe complications, including death.Given the lack of survival benefit provided by this device, the risk-benefit ratio and costs of using systematic endovascular cooling catheters instead of other methods should therefore be carefully appraised.Additionally, endovascular cooling catheters should be promptly removed at the end of the ttm period, and a new standard central venous catheter inserted if central venous access is still required, ideally in the subclavian vein.Whether the risk of thrombosis depends only on the catheter device or on the biological changes induced by hypothermia and cardiac arrest remains to be investigated.Cardiac arrest and thrombotic complication.This was a resuscitated post cardiac arrest (ca) patient.It is known that in such patient population blood flow is compromised and inflammatory status is high, which may lead to development of thrombotic complications.Disseminated vascular endothelium damages explain that this phenomenon of ischemia-reperfusion evolves toward systemic inflammation: production of cytokines (il-1, il-6, il-8, tnf-[?]), complement activation, synthesis of metabolites of arachidonic acid, expression of leukocyte adhesion molecules by endothelial cells are all stimuli for activation, and chemotaxis of polymorphonuclear neutrophils at the origin of the inflammatory response.Sequestration of activated neutrophils in the lungs and other organs is a major engine for development of multiple organ failure.It is plausible that infectious lesions of digestive origin are involved, in relation to bacterial translocation from altered gastrointestinal mucosa.It would explain the high levels of plasma endotoxin observed in the most critically ill patients.This activation of the systemic inflammatory response associates with changes in coagulation, generating secondary endothelial damage, in turn responsible for thrombosis and increased capillary permeability [adrie c, adib-conquy m, laurent i, monchi m, vinsonneau c, fitting c, fraisse f, dinh-xuan at, carli p, spaulding c, dhainaut jf, cavaillon jm.Successful cardiopulmonary resuscitation after cardiac arrest as a "sepsis-like" syndrome.Circulation.2002;106:562-568.} nicolas mongardon et all described that the prognosis for postcardiac arrest patients remains very bleak, not only because of anoxic-ischemic neurological damage, but also because of the "postcardiac arrest syndrome," a phenomenon often severe enough to cause death before any neurological evaluation.This syndrome includes all clinical and biological manifestations related to the phenomenon of global ischemia-reperfusion triggered by cardiac arrest and return of spontaneous circulation.The main component of the postcardiac arrest syndrome is an early but severe cardiocirculatory dysfunction that may lead to multiple organ failure and death.[postcardiac arrest syndrome: from immediate resuscitation to long-term outcome in ann intensive care.2011; 1: 45].Cardiovascular support relies on conventional medical and mechanical treatment of circulatory failure.Hemodynamic stabilization is a major objective to limit secondary brain insult.When the cause of cardiac arrest is related to myocardial infarction, percutaneous coronary revascularization is associated with improved prognosis; early angiographic exploration should then be discussed when there is no obvious extracardiac cause.Therapeutic hypothermia is now the cornerstone of postanoxic cerebral protection.Its widespread use is clearly recommended, with a favorable risk-benefit ratio in selected population.Neuroprotection also is based on the prevention of secondary cerebral damages, pending the results of ongoing therapeutic evaluations regarding the potential efficiency of new therapeutic drugs.The postcardiac arrest shock, initially described by negovsky in 1975, is a mixed shock, with cardiogenic and vasodilatory components [2] and is characterized by a severe but reversible systolic dysfunction.Left ventricular dysfunction usually begins early, within minutes after rosc, and is completely reversible within 48 to 72 hours.It takes the form of myocardial systolic and diastolic dysfunctions, even in the absence of coronary cause.Obviously, this has to be distinguished from the focal hypocontractility caused by a coronary thrombosis, which evolves on its own.The circulatory failure might also be favored by the occurrence of relative adrenal insufficiency, as observed during septic shock: clinical data show that approximately half of patients have relative adrenal insufficiency similar to what can be observed in septic shock.Among these patients, shock-related deaths seem to be more frequent in the group of those with adrenal insufficiency [12].[laurent i, monchi m, chiche jd, joly lm, spaulding c, bourgeois b, cariou a, rozenberg a, carli p, weber s, dhainaut jf.Reversible myocardial dysfunction in survivors of out-of-hospital cardiac arrest.J am coll cardiol.2002;40:2110-2116.[pubmed]; pene f, hyvernat h, mallet v, cariou a, carli p, spaulding c, dugue ma, mira jp.Prognostic value of relative adrenal insufficiency after out-of-hospital cardiac arrest.Intensive care med.2005;31:627-633.[pubmed].Asymptomatic thrombi in vein in studied post cardiac arrest patients, presence of which was reported in this article, could potentially lead to pe in such patient population.It is known that in turn, that in rare cases pe can lead to acute respiratory distress.Event assessed as possibly related to the catheter even in this predisposed patient.Refractory (unmanageable) cardiogenic shock.Cardiogenic shock is a life-threatening condition that occurs in response to reduced cardiac output in the presence of adequate intravascular volume and results in tissue hypoxia.Cardiogenic shock has several underlying etiologies, with the most common being acute myocardial infarction.Refractory cardiogenic shock presents as persistent tissue hypoperfusion despite administration of adequate doses of two vasoactive medications and treatment of the underlying etiology.Investigators of the shock trial reported a long-term mortality benefit of emergency revascularization for shock complicating ami.Since the publication of the shock trial and subsequent guideline recommendations, the increase in community-based use of percutaneous coronary intervention for this condition has resulted in a significant decline in mortality.Despite these successes in the past 15 years, mortality still remains exceptionally high, particularly in patients with refractory cardiogenic shock [reyentovich a, et al.Management of refractory cardiogenic shock.Nat rev cardiol.2016 aug;13(8):481-92].One of other conditions that may cause cardiogenic shock include pulmonary embolism (sudden blockage of an artery in the lung).It was not specified in the article, how late after occurrence of pe the patient experienced cardiogenic shock, if pe was resolved or not at that time, and if authors believe that cardiogenic shock was related to pe.It is also not specified when pe occurred in relation to cooling procedure and if this patient had defined by ultrasonography-confirmed thrombosis at the catheter insertion site.With that, based on available information, event of cardiogenic shock is assessed as possibly related to pe.Pe in term assessed as possibly related to asymptomatic thrombus i the patient's vein.At the same time, this patient was admitted after resuscitated cardiac arrest.Mortality rate is this population is high and it is possible that cardiogenic shock was related to the patient's underling clinical condition.
|