• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PENUMBRA, INC. PENUMBRA COIL 400; HCG, KRD

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PENUMBRA, INC. PENUMBRA COIL 400; HCG, KRD Back to Search Results
Catalog Number 4004C0513
Medical Device Problem Codes Failure to Advance (2524); Mechanical Jam (2983)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 01/18/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Results: the pet lock was intact on the proximal end of the pusher assembly.The pusher assembly was kinked in multiple locations.The embolization coil was detached from the pusher assembly and the stretch resistant wire was fractured (sr wire).Conclusions: evaluation of the returned pc400 revealed the sr wire was fractured and the embolization coil was detached.Fracture of the sr wire typically occurs due to forceful retraction of the pc400 against resistance, and will allow the embolization coil to detach.A demonstration pc400 was advanced and retracted through the lantern without issue and, therefore, the root cause of the reported resistance could not be determined.Penumbra coils and catheters are inspected during in-process inspection and during quality inspection after manufacturing.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.This report is associated with mfr report number: 3005168196-2018-00348.
 
Event or Problem Description
The patient was undergoing a coil embolization procedure in the right internal iliac artery (iia) using penumbra coil 400s (pc400s) and a lantern delivery microcatheter (lantern).During the procedure, the physician successfully deployed and detached five pc400s slightly proximal at the distal bifurcation of the right iia aneurysm using the lantern.While advancing a new pc400 through the lantern, the physician noticed that the pc400 was unable to fully advance out of the lantern and became stuck.Upon attempting to retract the pc400, it unintentional detached within the lantern.The lantern was therefore removed, and the physician pulled the pc400 out, and then reinserted the same lantern.Next, the physician attempted to advance a new pc400, however the pc400 was unable to advance and became stuck again.Upon attempting to retract the pc400 it unintentionally detached.The physician therefore removed the lantern with the pc400 inside, and completed the procedure using a new microcatheter and a new coil.There was no report of an adverse effect to the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PENUMBRA COIL 400
Common Device Name
HCG, KRD
Manufacturer (Section D)
PENUMBRA, INC.
one penumbra place
alameda CA 94502
MDR Report Key7277525
Report Number3005168196-2018-00347
Device Sequence Number2301329
Product Code HCG
UDI-Device Identifier00814548010861
UDI-Public00814548010861
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K120330
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source distributor,foreign,health pr
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 01/21/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Physician
Device Expiration Date04/08/2025
Device Catalogue Number4004C0513
Device Lot NumberF75487
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/31/2018
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/21/2018
Initial Report FDA Received Date02/16/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age82 YR
Patient Weight54
-
-