| Catalog Number |
2030-6540-1 |
| Medical Device Problem Codes |
Component Missing (2306); Packaging Problem (3007)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
01/22/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
An event regarding a packaging issue involving a bone screw was reported.The event was confirmed.Device evaluation and results: the device and its packaging were not returned.However, an image was provided showing the packaging box with a torn opening flap and an ifu, without the blister pack and the implant.No medical records were received for review with a clinical consultant.No adverse consequences to the patient were noted.Device history review: review indicated all devices were manufactured and accepted into final stock with no reported discrepancies.Complaint history review: there have been no other events for the lot referenced.The reported event was found to be a nonconformance.The potential root cause is more of a human error as the part and blister assembly are required to be visually inspected during the contents to box and sealing operation.This is an isolated occurrence.Not returned.
|
| |
|
Event or Problem Description
|
|
It was reported that the torx cancellous bone screw box was opened and found to contain no blister pack nor was there a screw.The container had the original plastic wrap on it.
|
| |
|