| Model Number |
1000 |
| Medical Device Problem Codes |
Mechanical Problem (1384); Low impedance (2285); Communication or Transmission Problem (2896)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
01/23/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Suspect device software version: version 1.0.2.2.
|
| |
|
Event or Problem Description
|
|
It was reported that a physician was experiencing communication issues when interrogating a patient's generator.The error code displayed was "error code 254 - apply changes interrupted".From review of the programming data, it was noted that the device was in a magnet inhibit state on the day of this office visit.The magnet inhibit state triggers when a magnet is placed over the generator to inhibit vns stimulation per user discretion.In response to attempting to program the generator in this state, the programmer interpreted the response with the "error code 254" message, which is a generic message for when a programming event may not have been successful.This message does not contain specific context to the issue, but implies that re-performing the programming operation might lead to success, although in this case, programming could not occur as long as a magnet was present over the device.It was noted that this "magnet inhibit" condition was not seen in any of the patient's previous sessions or with any other similar devices seen at other office visits.Follow up with the physician indicated that a magnet was not close to the generator at the time of programming.It was noted, however, that the patient's mother had a magnet taped over the patient's generator entering the clinic, and then removed it during the session.Further analysis was performed for this event of failure to program due to the magnet inhibition state.It was noted that prior to making any therapy programming changes to a generator, the programmer will send a command to end any active stimulation.In the presence of a magnet, this command is treated as an error by the programmer software, and is interpreted as a failure to end active stimulation.This causes the programmer to not proceed with the programming event, and triggers the generic "error code 254" message.At the patient's follow up visit, it was noted that the patient's magnet was not taped over his chest, and that the programming event was performed successfully with no issues.No additional relevant information was received to date.
|
| |
|
Event or Problem Description
|
|
During a follow up clinic visit ((b)(6) 2018), it was reported that the physician's assistant was still unable to program the generator, and was receiving the same error message "error code 254 - apply changes interrupted".Troubleshooting steps were performed to indicate that the physician's programming system was functioning as intended and was able to program other devices.Troubleshooting steps were also performed to account for electromagnetic interference, but the programming system was still unable to program the patient's device.After a generator reset was performed, it was noted that the device was able to be programmed to desired settings.Diagnostics, however, indicated low impedance.X-rays were reviewed for possible lead breaks, and no gross fractures were noted.Further review of tablet data received after this visit suspected that there may be a hardware malfunction with the generator's magnet detection circuitry, and that in this state, the device does not deliver any therapy despite diagnostic results indicating that current was delivered.Impedance values and diagnostics were indicated to be ok and within normal limits prior to (b)(6) 2018, when it was noted that the patient was admitted to the hospital for strep throat that was confirmed to be unrelated to vns.The lead impedance measured at this date was noted to still be within normal limits.The same impedance was noted throughout the (b)(6) 2018 office visit and for the beginning of the (b)(6) 2018 office visit, prior to the generator reset.This value was suspected to be a stale value taken from the (b)(6) 2018 impedance measurement.After the generator reset, the device was noted to have low impedance.All subsequent device interrogations and diagnostics reported the same low impedance value.At a later follow up clinic visit ((b)(6) 2018), diagnostics were performed again and still indicated low impedance.It was noted that the device failed to detect any magnet swiping, despite multiple attempts at swiping the magnet.The output current and sensing capabilities were disabled at the end of this visit.A review of the device history records for the generator indicated that the device met all specifications and passed all quality inspections prior to distribution.No surgery has occurred to date.No additional relevant information was received to date.
|
| |
|
Event or Problem Description
|
|
A generator replacement surgery occurred and the device was successful replaced, with indication that impedance of the new system was ok, within normal limits.The explanted device was returned for analysis.Analysis has not been completed to date.No further relevant information has been received to date.
|
| |
|
Event or Problem Description
|
|
Product analysis was completed for the returned generator.Visual examination was completed and found markings typical of surgical procedures.No other surface abnormalities were noted on this device.An interrogations test was performed and no impedance value was noted using a test resistor.The generator was programmed to deliver an output.No output was observed in normal mode or when the magnet was applied and removed.This indicated that the reed switch was stuck closed.The generator was opened for analysis.However the reed switch was operating properly after opening the generator.This may have been due to the vibration during the opening can process.The generator was subjected to and successfully completed a final electrical test and operated within specifications.In pa, it was observed that if a magnet was applied to the device for an extended period of time (such as overnight), the switch would remain stuck in a closed state for brief periods after the magnet was removed (between a few seconds and thirty seconds).After the battery was removed from the rest of the pcba, the same sticky behavior could not be repeated.No further relevant information has been received to date.
|
| |
|
Additional Manufacturer Narrative
|
|
Udi ¿ corrected data; incorrect udi number was reported on initial mdr.
|
| |
|