• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT MEDICAL OPTICS TECNIS 1-PIECE; MONOFOCAL IOLS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT MEDICAL OPTICS TECNIS 1-PIECE; MONOFOCAL IOLS Back to Search Results
Model Number ZCB00
Medical Device Problem Code Torn Material (3024)
Health Effect - Clinical Code No Information (3190)
Date of Event 01/18/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Unknown if lens was implanted.Attempts have been made to obtain missing information; however, to date, no response has been received.All pertinent information available to abbott medical optics has been submitted.
 
Event or Problem Description
It was reported that the inserter tore the intraocular lens during insertion into the eye.No additional information was provided to abbott medical optics.
 
Additional Manufacturer Narrative
Device evaluation: the product testing could not be performed as the product was not returned.The reported complaint cannot be confirmed.Manufacturing record review: the manufacturing process record was evaluated and no deviation was found during process related to the complaint issue reported.There was no discrepancy found during the mrr (manufacturing record review).The product was manufactured and released according to the specifications.A search revealed that no additional complaints for this production order have been received.Labeling review: the directions for use (dfu) was reviewed.The dfu adequately provides instructions and precautions for the proper use and handling of the product.Conclusion: based on the investigation results, there is no indication of a product quality deficiency.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TECNIS 1-PIECE
Common Device Name
MONOFOCAL IOLS
Manufacturer (Section D)
ABBOTT MEDICAL OPTICS
santa ana CA
MDR Report Key7278969
Report Number2648035-2018-00256
Device Sequence Number14047928
Product Code HQL
UDI-Device Identifier05050474531826
UDI-Public(01)05050474531826(17)211017
Combination Product (Y/N)N
Initial Reporter StateGA
Initial Reporter CountryUS
PMA/510(K) Number
P980040
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 06/19/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Physician
Device Expiration Date10/17/2021
Device Model NumberZCB00
Device Catalogue NumberZCB0000255
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 01/19/2018
Supplement Date Received by Manufacturer05/23/2018
Initial Report FDA Received Date02/16/2018
Supplement Report FDA Received Date06/19/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-