MAUDE Adverse Event Report: ABBOTT MEDICAL OPTICS TECNIS; MONOFOCAL IOLS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT MEDICAL OPTICS TECNIS; MONOFOCAL IOLS Back to Search Results
Model Number ZA9003
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Code Available (3191)
Date of Event 01/08/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Device evaluation: the product was not returned to the manufacturing site.Therefore, product testing could not be performed and the customer¿s reported complaint could not be verified.Manufacturing records review: the manufacturing records for the product was reviewed.The product was manufactured and released according to specification.A search revealed no additional investigation requests for this production order number.Labeling review: the directions for use (dfu) was reviewed.The dfu adequately provides instructions and precautions along with warnings for the proper use and handling of the device.As a result of the investigation there is no indication of a product quality deficiency and the reported issue could not be verified.The lens was inserted and removed.(b)(4).All pertinent information available to abbott medical optics has been submitted.
 
Event or Problem Description
It was reported that the haptic of a za9003 17.0 diopter intraocular lens (iol) was noticed broken after the lens was implanted.Therefore, it was removed and replaced during the same surgical procedure.An incision enlargement was made and sutures were used.There was no patient injury.Reportedly, the replacement lens was the same model.No additional information was provided.
 
Additional Manufacturer Narrative
Device available for evaluation? yes.Returned to manufacturer on: 04/12/2018.Device returned to manufacturer? yes.Device evaluation: the intraocular lens (iol) was returned to the manufacturer.Visual inspection using microscope magnification was performed.Loose fibers/particles were observed on lens related to the handling of the unit out of a sterile environment.Residues of lubricant material were also observed on lens.Both haptics were observed detached.The detached haptic pieces were not returned.The condition in which the sample returned is consistent with a product that was handled and prepared for a surgical process.The complaint issue reported as haptic damaged was not verified.However, haptic detached was identified on sample returned.Based on the analyzed of the returned, there is no indication of product quality deficiency.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TECNIS
Common Device Name
MONOFOCAL IOLS
Manufacturer (Section D)
ABBOTT MEDICAL OPTICS
santa ana CA 92705
MDR Report Key7279088
Report Number2648035-2018-00258
Device Sequence Number2248338
Product Code HQL
UDI-Device Identifier05050474528956
UDI-Public(01)05050474528956(17)220728
Combination Product (Y/N)N
Initial Reporter StateMN
Initial Reporter CountryUS
PMA/510(K) Number
P980040
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source distributor,health profession
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 04/26/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Physician
Device Expiration Date07/28/2022
Device Model NumberZA9003
Device Catalogue NumberZA90030170
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/12/2018
Initial Date Received by Manufacturer 01/22/2018
Supplement Date Received by Manufacturer04/16/2018
Initial Report FDA Received Date02/16/2018
Supplement Report FDA Received Date04/27/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age68 YR
-
-