|
SMITH & NEPHEW, INC. LEGION HK 15MM BOLT - SLEEVE; PROSTHESIS, KNEE, FEMOROTIBIAL, CONSTRAINED, CEMENTED, METAL/POLYMER
|
Back to Search Results |
|
| Catalog Number |
71421387 |
| Medical Device Problem Code |
Mechanical Problem (1384)
|
| Health Effect - Clinical Code |
Joint Disorder (2373)
|
| Date of Event |
01/30/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
Revision of insert and legion hinge mechanism due to instability of de knee.Patient was walking and suddenly the knee felt unstable and she wasn¿t able to walk further.She didn¿t fall or injured herself in any way.We opened the knee and found some metalized tissue.Removing the bolt was very easy.We weren¿t able to remove the sleeve without first detaching the hinge mechanism.So we removed the mechanism according the surgical technique.Because the sleeve was still in the lgn hk link assy sz 2-5 fem, we weren¿t able to remove the lgn hk link assy sz 2-5 fem.Finally we turned the lgn hk link assy sz 2-5 fem 180 degrees and unscrewed the insert.With some force we managed to get the lgn hk link assy sz 2-5 fem and sleeve out of the tibia.We saw some black residue on the tibia base at the position of the insert screw.It looked like metalized residue as well.
|
| |
|
Additional Manufacturer Narrative
|
|
Results of investigation: the associated complaint devices were returned and evaluated.A visual examination indicated that there were signs of scratching on several of the as-received components that may have been due to the reported instability.The remaining hinge components were not received; therefore further analysis could not be conducted.The immediate cause of the instability could not be determined.No material or manufacturing deviations were observed in this investigation.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.A clinical analysis indicated based on the information available, the root cause of the ¿instability¿ and reported scratches cannot be concluded.No report as to the patient¿s current status has been submitted.Based on this investigation, the need for corrective action is not indicated.No further investigation is warranted for this complaint; however we will continue to monitor for future complaints and investigate as necessary.Should additional information be received, the complaint will be reopened.We consider this investigation closed.
|
| |
|
Search Alerts/Recalls
|
|
|