SMITH & NEPHEW, INC. BIOLOX DELTA HEAD 32 MM 12/14 S / +0; PRSTHSIS,HIP,SMI-CONSTRAIND,METAL/CRAMIC/POLYMRCMNTDORNON-POROUS,UNCMNTD
|
Back to Search Results |
|
| Catalog Number |
76539160 |
| Medical Device Problem Code |
Insufficient Device Problem Information (3190)
|
| Health Effect - Clinical Code |
Pain (1994)
|
| Date of Event |
01/24/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that a surgical intervention was performed due to thigh pain and thickening of lateral femoral cortex.The liner, the stem and the head were removed.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: the associated complaint device was not returned for evaluation.Our clinical investigation noted that no patient information has been submitted for review.All documents and images provided as of this date have been reviewed and considered and unless noted do not contribute to the clinical investigation.Based on the limited information, the root cause of the ¿thigh pain and thickening of lateral femoral cortex at femoral implant tip¿ cannot be concluded but cortical thickening can be a result of micro motion.No information has been forthcoming regarding the current status of this patient.Should any relevant clinical/medical information become available this complaint can be re-assessed.A review of the manufacturing records for the listed batch did not reveal any deviation from the standard manufacturing processes.Without the return of the actual product involved, our investigation of this report is inconclusive.If the device or new information is received in the future, this complaint can be re-opened.No further actions are being taken at this time.We consider this investigation closed.
|
| |
|
Search Alerts/Recalls
|
|
|