MEDTRONIC IRELAND RESOLUTE ONYX RX; STENT, CORONARY, DRUG-ELUTING Back to Search Results
Catalog Number RONYX22530UX
Medical Device Problem Code Material Deformation (2976)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 01/16/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
It was reported that the physician intended to use a resolute onyx drug eluting stent to treat a lesion located at the distal right coronary artery which had moderate tortuosity and mild calcification.There were no abnormalities reported in relation to anatomy.There was no damage noted to packaging and no issues noted when removing the device from the hoop/tray.The device was inspected with no issues identified.The lesion was pre-dilated.Resistance was encountered when advancing the device and excessive force was not used during delivery.It was reported that the stent deformation occurred in-vivo during positioning/advancement.It was reported that the stent struts were damaged.It was informed that the physician complete the procedure using a 2.5x18 non-medtronic device.No patient injuries were reported.
 
Additional Manufacturer Narrative
Evaluation summary: protective sheath and stylette were returned with device.Numerous kinks were evident along the hypotube.The stent was positioned on the balloon between the marker bands as per specifications.Deformation was evident to the 11-14th and 22nd stent wraps, with struts raised and bunched.No deformation was evident to the distal tip.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RESOLUTE ONYX RX
Common Device Name
STENT, CORONARY, DRUG-ELUTING
Manufacturer (Section D)
MEDTRONIC IRELAND
parkmore business park west
galway
MDR Report Key7279649
Report Number9612164-2018-00370
Device Sequence Number1066502
Product Code NIQ
UDI-Device Identifier00643169556751
UDI-Public00643169556751
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
P160043
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 05/25/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/07/2019
Device Catalogue NumberRONYX22530UX
Device Lot Number0008524028
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/20/2018
Initial Date Received by Manufacturer 01/24/2018
Supplement Date Received by Manufacturer05/08/2018
Initial Report FDA Received Date02/19/2018
Supplement Report FDA Received Date05/25/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-